Scope Calibration Option
Summary of Commands and Queries
19
Table 2. Overlapped and Coupled Commands
Command Overlapped Coupled
SCOPE(?) Yes
Yes
TRIG(?) Yes
No
OUT_IMP(?) Yes
Yes
SCOPE(?) <value >
Description
Programs the 5080A to use the Megohm Option, if installed.
Parameters
<value> = OFF
Turns the Scopt option off. Programs 0 V, 0 Hz
output at the NORMAL terminals.
VOLT
Oscilloscope ac and dc voltage mode. Programs
20 mV p-p, 1 kHz, output at the SCOPE BNC,
output impedance 1 M
Ω
standby if from OFF or
previously in standby.
EDGE
Oscilloscope edge mode. Programs 25 mV p-p,
1 MHz, output at the SCOPE BNC, standby if
from OFF or previously in standby.
LEVSINE Oscilloscope-leveled sine mode. Programs
30 mV p-p, 50 kHz, output at the SCOPE BNC,
standby if from OFF or previously in standby.
MARKER Oscilloscope Marker mode. Programs the period
to 1 ms, output at the SCOPE BNC, standby if
from OFF or previously in standby.
Example
SCOPE VOLT; OUT -2 V, 0 Hz
(dc voltage, -2 V)
SCOPE VOLT; OUT 4 V, 1 kHz
(ac voltage, 4 V p-p, 1 kHz)
SCOPE EDGE; OUT 0.5 V, 5 kHz
(Edge, 0.5 V p-p, 5 kHz)
SCOPE LEVSINE; OUT 1 V, 200 kHz
(Leveled sinewave, 1 V p-p,
200 kHz)
SCOPE MARKER; OUT 2 MS
(Marker period of 2 ms)
Query
SCOPE?
Returns the current mode of operation (OFF,
VOLT, EDGE, LEVSINE, or MARKER).
TRIG(?) <value >
Description
Programs the scope options trigger out BNC.
Parameters
<value> = OFF
Turns the trigger output off.
DIV1
Turns the trigger output on. Frequency is the
same as the signal at SCOPE output.
DIV10
Turns the trigger output on. Frequency is 1/10
the signal at SCOPE output.
DIV100
Turns the trigger output on. Frequency is 1/100
the signal at SCOPE output.
Example
TRIG OFF
Turns the trigger output off.
Query
TRIG?
Returns the mode of the Megohm option (OFF,
DIV1, DIV10, or DIV100).