Scope Calibration Option
Verification Tables
21
Table 3. Voltage Function Verification: AC Voltage into a 1 M
Ω
ohm Load (cont.)
Nominal Value (p-p)
Frequency
Measured Value (p-p)
Deviation (mV)
1-Year
Spec. (mV)
20.0 V
1 kHz
70.2
50.0 V
45 Hz
175.2
50.0 V
100 Hz
175.2
50.0 V
1 kHz
175.2
105.0 V
100 Hz
6050
105.0 V
1 kHz
6050
Table 4. Voltage Function Verification: AC Voltage into a 50
Ω
Load
Nominal Value (p-p)
Frequency
Measured Value (p-p)
Deviation (mV)
1-Year
Spec. (mV)
5.0 mV
45 Hz
0.2175
5.0 mV
100 Hz
0.2175
5.0 mV
1 kHz
0.2175
10.0 mV
100 Hz
0.235
10.0 mV
1 kHz
0.235
20.0 mV
1 kHz
0.375
44.9 mV
45 Hz
0.35715
44.9 mV
1 kHz
0.35715
50.0 mV
1 kHz
0.375
100.0 mV
100 Hz
0.55
100.0 mV
1 kHz
0.55
200.0 mV
1 kHz
0.9
449.0 mV
45 Hz
0.7715
449.0 mV
1 kHz
0.7715
500.0 mV
1 kHz
1.95
1.0 V
100 Hz
3.7
1.0 V
1 kHz
3.7
2.0 V
45 Hz
7.2
2.0 V
100 Hz
7.2
2.0 V
1 kHz
7.2