Chapter 4: Maintenance
Cleaning Summary
88
CyTOF XT User Guide
Cleaning Summary
Table 5. Cleaning summary
Part
Frequency and Cleaning Agent
Procedure
Nebulizer
Daily
Contrad 100/Decon 90 (dilute to 10% in DIW)
1
Draw 10% Contrad/Decon through side arm
and sample inlet.
2
Soak in 10% Contrad/Decon overnight
(16
–
24 hr).
3
Rinse with DIW and soak in DIW until next
use.
IMPORTANT
To prevent damage:
•
Do not sonicate.
•
Do not clean with acidic detergent (for
example, Citranox).
•
Do not soak in Contrad for longer than 24 hr.
Nebulizer rest
Weekly
Type I ultrapure (18.2MΩ) water (DIW)
Replace the DIW within the nebulizer rest tube
(15 mL capacity).
Spray chamber
Weekly or after 115 hr of use
Isopropanol, Kimwipes
1
Clean with isopropanol and use Kimwipes
to remove residue.
2
Air-dry thoroughly before reinstalling on
instrument.
Injector
Weekly or more frequently as needed*
* Check the injector every 50 hr of sample
acquisition. Clean injector if buildup or
discoloration is observed at tip.
10% Contrad 100/Decon 90 in DIW
1
Soak in 10% Contrad/Decon for 1 hr.
2
Scrub with brush.
3
Rinse with DIW.
4
Dry thoroughly before reinstalling on
instrument.
Interface cones
(sampler cone and
skimmer-reducer
assembly)
Weekly or after 115 hr of use
10% Citranox in DIW, Bransonic M1800
Ultrasonic Bath (or similar sonicator)
1
Place cones with adapters into the cleaning
container.
2
Sonicate in 10% Citranox for 15 min. Do not
soak for longer than 15 min.
3
Rinse cones, adapters, and container with
DIW. Sonicate in DIW for 15 min.
4
Repeat step 3.
5
Air-dry thoroughly before reinstalling.
Autosampler fluidic
lines and nebulizer
Weekly or after 115 hr of use
10% bleach solution
(diluted to 10% v/v in DIW)
1
Turn off the nebulizer gas.
2
Place the nebulizer in the nebulizer rest
and replace the nebulizer rest tube with an
empty tube labeled Bleach Waste.
3
Run 10% bleach solution through the
Autosampler fluidic lines and nebulizer.
4
Run DIW through the Autosampler fluidic
lines and nebulizer.