Chapter 4: Maintenance
Remove and Clean the Torch Assembly
CyTOF XT User Guide
117
2
Pull the torch assembly straight toward you to remove it.
CAUTION
Finger cut hazard. Broken glass may cause injury or cutting of
fingers or hands. To avoid breakage, keep the torch assembly straight while
removing and replacing.
3
Carefully turn the white retaining ring counterclockwise to remove the ring.
CAUTION
Finger cut hazard. Broken glass may cause injury to fingers or
hands.