Chapter 4: Maintenance
Clean the Nebulizer
58
CyTOF XT User Guide
Cleaning Method for Manual Clog Removal: Hot DIW
NOTE
This method should be utilized to remove stubborn clogs when the daily cleaning
method is unsuccessful.
Materials
Table 4. Materials for nebulizer cleaning (Hot DIW)
Item
Source
Heat-resistant glove or mitt
Major laboratory supplier (MLS)
Beaker
MLS
Nebulizer-to-aspirator tubing:
3/16-inch OD, 1/8-inch ID
1/8-inch OD, 1/16-inch ID
Nebulizer Cleaning Kit
Type I ultrapure, >18.2 MΩ (DIW)
MLS
Microwave
MLS
Vacuum aspirator
User-supplied
Flashlight
Fluidigm supplied
Light microscope (optional)
MLS
Procedure
IMPORTANT
Do not sonicate the nebulizer or put pressure on the fragile nebulizer tip.
1
Fill a small beaker halfway with DIW.
2
Boil the DIW.
3
Using a heat-resistant glove or mitt, carefully transfer the beaker to a lab bench.
4
Connect a vacuum aspirator (suitable for liquid waste disposal) to the nebulizer using
the appropriate tubing.
•
If the clog is near the nebulizer tip, connect the tip to the aspirator using 3/16inch
outside diameter (OD), 1/8-inch inside diameter (ID) tubing.
•
If the clog is near the sample inlet, connect the sample inlet to the aspirator using
1/8-inch OD x 1/16-inch ID tubing.
5
Submerge the open end of the nebulizer (tip or sample inlet) and side arm fitting
(nebulizer gas inlet) of the nebulizer in the hot DIW.
6
Draw DIW through the nebulizer until the clog is removed.
WARNING
Hot surface hazard. The nebulizer may be very hot. Carefully
remove the nebulizer from the beaker using a heat-resistant glove or mitt.
7
(Optional) Visually inspect the nebulizer using a light microscope to verify that the clog
has been removed.
8
Check the nebulizer spray.
a
Reassemble the nebulizer: Connect the nebulizer gas line and the nebulizer line to
the nebulizer assembly.