Chapter 4: Maintenance
Clean the Vacuum Interface Cones
CyTOF XT User Guide
71
c
Add the top adapter.
d
Place the skimmer-reducer assembly on the top adapter with the screws facing up.
2
Sonicate cones in 10% Citranox. Do not soak for longer than 15 min.
3
Pour off the Citranox and rinse the cones, adapters, and container with DIW. Then stack
cones as follows:
IMPORTANT
Add DIW as each part is placed in the container to ensure that all parts
contact the DIW.
a
Stack the bottom adapter first.
b
Place the sampler cone face down on the bottom adapter.
c
Add the top adapter.