Chapter 4: Maintenance
Autosampler Module Maintenance
106
CyTOF XT User Guide
11
Position the new syringe plunger to the center of the stroke.
12
Insert the luer end of the new syringe into the valve and rotate it until finger tight.
13
Pull the plunger down until it reaches the syringe drive stem.
14
Position the plunger so that the plunger button sets into the channel of the drive stem.