Chapter 4: Maintenance
Nebulizer Assembly Maintenance
112
CyTOF XT User Guide
2
Hold the nebulizer mount with one hand. With the other thumb on the flat surface of the
spring assembly, gently turn the pin to align it with the slot to release the spring
assembly from the nebulizer mount.
3
Remove the spring assembly from the nebulizer mount. Clean it by wiping with a
Kimwipe moistened with isopropanol to remove any residue.