Figure 16. CyTOF XT flow diagram of the sample injection process
Acquisition of Pelleted Samples
The Autosampler can resuspend pelleted samples with Maxpar CAS Plus and add EQ beads
prior to acquisition. You can specify a dilution volume between 150 and 4,000 µL. The
recommended final concentration is no greater than 1 million cells/mL.
IMPORTANT
We recommend keeping the samples as pellets to avoid having samples in
suspension for extended periods of time. Only 5 mL tubes should be used for pelleted
samples. Pelleted samples in 15 mL skirted conical tubes are not part of the recommended
workflow.
Acquisition of Suspension Samples
You can choose to manually resuspend samples with Maxpar CAS Plus and EQ beads prior
to acquisition and place pre-diluted cells in the carousel. The recommended final
concentration is no greater than 1 million cells/mL. Samples should not be kept in
suspension for any longer than 4 hr prior to starting acquisition of that sample. Batch
acquisitions should be set up such that suspension samples are acquired first.
Large suspension samples can be acquired from 15 mL skirted conical tubes. When running
a 15 mL tube as part of a batch, it should be the first sample to be acquired.