Chapter 4: Maintenance
Reassemble the Injector and Spray Chamber
76
CyTOF XT User Guide
4
Place the heater assembly into its holder on the left side of the instrument. Install the
heater assembly cap on the heater assembly and lock the cap in place by gently
pushing down and then turning clockwise.
5
Make sure the injector is completely dry.
6
With the injector cap off, insert the injector into the injector holder until it reaches a hard
stop.
IMPORTANT
If the injector tip is chipped or broken, do not use. Replace the injector.