Chapter 4: Maintenance
Clean the Injector and Spray Chamber
CyTOF XT User Guide
65
2
Carefully remove the heater assembly cap, keeping it in the upright position because
the spray chamber and the injector holder may not be secured tightly to the heater
assembly cap.
3
Remove the spray chamber and injector holder from the heater assembly cap.
IMPORTANT
Carefully remove the spray chamber since the injector holder may loosen
from the heater cap and fall out.