Operating Manual for Operating Table SU-02.0

_____________________________________________________________________________________________________

1.

Proper use and application 

1.1.

Application

Operating   table   SU-02  is   designed   for   supporting   the   patient   during   carrying   out   treatments   and 
operations.  The table with proper accompanying equipment can be used for procedures of general 
surgery,   vascular   surgery,   cardiosurgery,   neurosurgery,   urology,   gynecology,   proctology,   ENT, 
endoscopy, laparoscopy, ophthalmology, trauma surgery, oncology surgery, dental surgery, plastic 
surgery and others. Segments of the table-top are radiotransparent. 

1.2.

General requirements

The product is intended to be used indoors. Required climatic conditions: temperature from +10 to 
+40ºC,   acceptable   change   of   surrounding   temperature   during   8   hours   should   not   exceed   20°C, 
relative humidity of the air should range from 30 to 80%, atmospheric pressure from 700 to 1060 hPa. 
The product should be used, maintained and serviced according to the instructions contained in this 
manual.

Caution!

Using, maintaining and servicing the product in other way than indicated in this 
manual is not permitted and may lead to damages for which the user is to blame 
and for which the producer is not responsible. 

Installation   of   other   accessories   than   those   offered   by   the   producer   for   the 
product is allowed only on the basis of a written acceptance of the producer. 

1.3.

Duties of the user 

User:

  any individual or corporate body who uses the product  as its owner, lessee, pledgee or who  

has a different right to the product as well as an entity who uses the product on its own or on whose  
behalf it is used. 

The user must ensure that the product shall be used exclusively in conformity with its destination and  
that it is used in appropriate conditions and in consistence with this manual. The user is also obliged to  
take all necessary precautions in order to prevent all life and health hazards concerning the user, 
patients   and   any   third   party.   Only   authorised   persons   who   underwent   special   training   and   are 
acquainted with this manual may operate the product. The user must also ensure that all persons who 
operate the product have read, understood and apply instructions contained in this manual. 

1.4 Description of the product

The table consist of table base and table top. 
The table has a mobile base (standard or Supermobile – with large wheel), equipped with central 
blockade, which is released with a foot lever. The top of the operation table is permanently attached to 
the base. The table-top consists of segments, and their number and configuration is deliver according 
to a customer order.  The table top is X-ray permeable. The table top is provided with polyurethane 
mattresses.  On both sides of the table top there are side strips which allow to install accessories. 
Adjustments of the position of individual segments, as well as column movements are performed by 
mechanical systems. 
Thanks to the wide range of offered accessories the table can be used for different types of operations 
according to client wish. List of accessories can be found in Paragraph 4 - Installation and operation of 
accessories.

The producer reserves the right to introduce in the product structural modifications resulting 
from technical progress which are not covered in this user manual.

 

The   producer   reserve   that   all   parameters   and   accessories   can   be   modified  or   change 
especially   construction,   technology   and   materials,   not   lowering   accepted   parameter 
technically-user and safeties of products.

_________________________________________________________________________________

page 6

Содержание SU-02

Страница 1: ...USER MANUAL Operating table SU 02 Factory no Version 12 00 ywiecka Fabryka Sprz tu Szpitalnego FAMED S A Appendixes 2 3 A List of Spare Parts...

Страница 2: ...edure was carried out in accordance with appendix VII of the directive Manufacturer ywiec Hospital Equipment Company FAMED S A Fabryczna 1 34 300 ywiec Phone numbers Infoline 48 33 866 63 75 24 hours...

Страница 3: ...er Every repair of the product must be done by a factory or an authorized service and recorded on the list of repairs which is supplied with the guarantee certificate Disregarding this requirement wil...

Страница 4: ...g the table avoid collisions Do not roll the table over electric cables If it is necessary to replace oil dangerous waste one should follow existing environmental protection regulations Notes concerni...

Страница 5: ...CHANGE OF ANGULAR POSITION OF THE FOOTRESTS 14 3 7 CHANGE OF ANGULAR POSITION OF DIVIDED FOOTRESTS 15 3 8 ROTATION OF THE FOOTREST 15 3 9 INSTALLATION AND REMOVAL OF THE FOOTREST 15 3 10 INSTALLATION...

Страница 6: ...exclusively in conformity with its destination and that it is used in appropriate conditions and in consistence with this manual The user is also obliged to take all necessary precautions in order to...

Страница 7: ...________________________________________________________________ 1 5Description of elements drawing 1 Operating table SU 02 standard version drawing 2 Operating table SU 02 Supermobile version _______...

Страница 8: ...angle change lever 17 Blockade of lever of back support angle adjustment 18 18 1 18 2 18 3 Kidney bench crank storage position adjustment position of kidney support adjustment position of lateral tilt...

Страница 9: ...m the floor 1000 mm 15 mm Backrest raise angle 70 Backrest lowering angle 40 Kidney bench raise angle 110 Footrest raise angle 4 and 5 sectional table top 15 Footrest raise angle 6 sectional table top...

Страница 10: ..._________________________________________________________________________ 1 7Use parameters The figure below indicates the most important functional parameters of the table drawing 4 _________________...

Страница 11: ...losely follow operating instruction for that equipment Improper operation may become a source of dangerous accidents There is a danger of serious burning of the patient through the contact with metal...

Страница 12: ...ng with the product manufacturer in order to ensure safe transport 2 2Unpacking and first use Caution If the product is installed by an authorised service of FAMED S A the user is released from the ob...

Страница 13: ...oes not additional operation or maintenance 3 2Longitudinal tilt of the table top to adjust longitudinal tilt of the table top trendelenburg and anti trendelenburg you should stand behind the backrest...

Страница 14: ...backrest segment Angle adjustment of the position of the back support segment Fig 1 3 position 2 is performed by muscle force with a gas spring back up In order to adjust the segment position push the...

Страница 15: ...be correctly meshed When the knob is being turned down one should check if the toothed bars are meshed 3 9Installation and removal of the footrest The footrest may be placed and removed from the seat...

Страница 16: ...he table The table in this version is equipped with central blockade of every wheel Pedal of central blockade has three positions upper direction blockade Fig 2 position 24 3 middle all wheels unblock...

Страница 17: ...ation of potential equalization in the operations room The cable is provided in standard table accessories 3 15 Collisions In some extreme positions of the table in particular when accessories install...

Страница 18: ...19 6 18 Specialist headrest WS 21 5 19 Laryngology ophthalmic headrest with track adjustment WS 21 7 20 Head rest for cervical kerb WS 21 8 21 Hand support WS 22 5 22 Body strap WS 23 0 23 gynaecologi...

Страница 19: ...e in any direction should not cause tilting or moving 2 Check mechanisms of adjustment of location of the segments controlled by gas springs by a change of position and the lock see item 3 Operation T...

Страница 20: ...roduct accessories should be reported immediately to a person in charge of such issues The bed which can not be safely operated e g damaged electric or mechanical elements must not be used till it is...

Страница 21: ...protections on pivot bolts Check whether the lock and table moving system work Check the hydraulic pump Check whether gas springs work correctly Table top check whether the trendelenburg mechanism wo...

Страница 22: ...ntification When sending asking any questions concerning the table and when ordering spare parts please give the serial number of the table placed on the nameplate and the guarantee certificate The se...

Страница 23: ...rating Table SU 02 0 _____________________________________________________________________________________________________ 7 2 Labels __________________________________________________________________...

Страница 24: ...ing and disinfection you should not use agents containing active oxygen and chlorine 5 Maximum authorized workload 150 kg 6 Attention read user manual 7 Longitudinal tilt Trendelenburg anti Trendelenb...

Страница 25: ...____________ 13 Table top height adjustment 14 Table s brake 15 Central wheel lock Notice The producer reserves the right to introduce in the product modifications resulting from technical progress wh...

Страница 26: ...C047310000000000 3 14 Potential balance cable 1 S1115000004258 2 15 Pin of potential balance 1 S11155000004256 2 16 Sliding sleeve 4 S30250100000200 2 17 Guiding sleeve 4 C073600000014000 2 18 Trende...

Страница 27: ...k rest rail I 2 C047311000020000 2 47 Crank set 1 C067204000000000 2 48 Protection 1 C067200000005000 3 49 Arm 1 C080300000700000 2 50 Screw 1 C080300000011000 2 51 Hand II 1 C047311000011000 2 52 Pro...

Страница 28: ...list Symbol SU 02 Edition 06 04 2007 Package should contain Operating table SU 02 set without headrest 1 pce and Headrest 1 pce SG 10 0 Potential equalization cord 1 pce index S11155000015625 Additio...

Отзывы: