Operating Manual for Operating Table SU-02.0
_____________________________________________________________________________________________________
segment, that is, by the handle which moves the mechanism of the gas spring. When the lever is
pressed, the required angle of the headrest may be chosen. Its release locks the headrest in the
chosen position.
Caution!
Make sure that the headrest is locked!
3.11 Installation and dismounting of mattresses
All mattresses can be removed without any tools. They are fixed on fixing pegs.
3.12 Table mobility
The table may move in all directions thanks to wheels fastened on its base. The table is provided with
movement blockade which is used during operations and procedures performed on the table.
Standard version and version with control panel on the head side
The table in this version is equipped with two feet, lowering of the feet cause the wheels to elevate and
loose contact with the floor, so that the table can not be rolled. Screwing or unscrewing of the feet
allows adjusting their projection. In order to lower the feet push the pedal toward the floor until you feel
resistance (Fig. 1,3, position 24.2). In order to elevate the feet lift the pedal to the upright position (Fig.
1,3, position 24.1).
Supermobile version of the table
The table in this version is equipped with central blockade of every wheel.
Pedal of central blockade has three positions:
-upper –direction blockade (Fig. 2, position 24.3),
- middle- all wheels unblocked (Fig. 2, position 24.1),
- lower – all wheels blocked (Fig. 2, position 24.2).
In “direction blockade” position one of the wheels is set for straight rolling. This position is used to roll
the table through long, straight distances, it facilitates turning the table. All-wheels blockade make the
table immobilized
Caution!
Throughout surgical procedures table wheels should be blocked.
The floor under the table must be free from any obstacles!
When rolling the table avoid collisions!
Do not roll the table over electric cables!
3.13 Antistatic properties
Table construction allows grounding of static charge through the following routes:
–
By antistatic wheels to conducting floor,
–
By potential equalisation clamp.
Operating table SU-02 should be used on antistatic floor. In case there is no such floor, a static charge
is grounded through the cord for potentials equalization. The potential equalising conductor is a
standard table accessory.
Antistatic properties of the table shall be maintained if mattresses produced by FAMED S.A. are used.
3.14 Potential equalising clamp
SU-02 operating table have inside system of potential equalization leading to the clamp (fig. 1-3, item
9) marked by symbol
.
_________________________________________________________________________________
page 16