Operating Manual for Operating Table SU-02.0
_____________________________________________________________________________________________________
1.8 Safety
The structure of the table assures its safe operation and use if the rules provided in this manual are
followed.
Caution!
The sign shown below says: “Caution – pay special attention to the Operating Manual”.
A label showing this sign is placed on any parts or mechanisms, which may prove to be
harmful to the patient or the personnel if their operation does not comply with the descriptions
found in this Operating Manual.
When operating the table one should pay attention to elements and mechanisms with this
label.
•
Throughout surgical procedures table wheels should be blocked.
•
When using the table close to medical equipment working on high frequencies and
defibrillators one should closely follow operating instruction for that equipment. Improper
operation may become a source of dangerous accidents. There is a danger of serious
burning of the patient through the contact with metal parts of the table or its equipment.
•
Throughout surgical procedures the table must be connected to the installation of potential
equalization in the operations room.
•
When locking a set position of the footrest, pay attention that the toothed bars are
correctly meshed.
•
When performing longitudinal (Trendelenburg or reverse Trendelenburg) and lateral tilting,
patient should be secured against uncontrolled sliding down from the operating table. Use
shoulder supports, supporting rollers, side supports, belts and grips as securing elements.
1.9 Critical parameters
Maximum work load
- 150 kg
Maximum authorized segment load
-head rest
10 kg
-back rest segment
55 kg
-seat segment
55 kg
- foot rest (simple)
15 kg
- foots rests (both load at the same time)
30 kg
fig.5
_________________________________________________________________________________