Operating Manual for Operating Table SU-02.0
_____________________________________________________________________________________________________
.
6.5 Control of technical condition
In order to ensure safe and proper technical condition of the product during the whole life of
the product, the product should undergo periodical technical inspections to be carried out by
the producer, an authorised service or authorised and trained technical staff of the customer
(at the expense of the user).
Only a positive result of product inspection can be the basis for its further operation.
The required service and its frequency are shown in the table below.
Subject of inspection
frequency
-
Checking functionality and general technical condition
(according to service instructions)
Every 12 months
Caution!
Every repair of the product should be recorded on the list of repairs enclosed to
the guarantee card on pain of guarantee loss!
6.5.
Control of technical condition
In order to provide safe and correct operation during product life, the user should perform the
following activities every six months:
Table basis
−
Check screw joints and protections on pivot bolts
−
Check whether the lock and table moving system work
−
Check the hydraulic pump
−
Check whether gas springs work correctly
Table top
−
check whether the trendelenburg mechanism works correctly
−
check whether the blockade of footrest fixing clamp works properly
−
check whether gas springs work correctly
Accessories
−
check functionality
−
check whether blockades work correctly
Potential equalising
cable
−
check the condition of insulation (cracks)
−
check how good the connection is
A checkout should involve a visual inspection and the noticed malfunctions should be handled as
provided in
item 6.6 Location of defects and their removal
.
6.6.
Location of defects and their removal
A list of repairs which may be carried out by the client on his own.
damage
Possible cause
removal
The table
cannot be
moved
locked wheels
Unlock blockade pedal
The
trendelenburg
mechanism
does not work
The mechanism of
trendelenburg wire
wrongly adjusted
Regulate wire mechanism in accordance with the
item: Table repairs and adjustment
The table is not
stable
Wrong adjustment
of table feet
Unscrew or screw down feet with a wrench
Plays in screw
joints
Plays appeared
during product
operation
Screw down with generally available wrenches
_________________________________________________________________________________
page 21