6.5 Control of technical condition

In order to ensure safe and proper technical condition of the product during the whole life of 
the product, the product should undergo periodical technical inspections to be carried out by 
the producer, an authorised service or authorised and trained technical staff of the customer 
(at the expense of the user).

Only a positive result of product inspection can be the basis for its further operation.  

The required service and its frequency are shown in the table below.

Subject of inspection

frequency

-

Checking functionality and general technical condition 
(according to service instructions) 

Every 12 months 

Caution!

 

Every repair of the product should be recorded on the list of repairs enclosed to 

the guarantee card on pain of guarantee loss!

6.5.

Control of technical condition

In   order   to   provide   safe   and   correct   operation     during   product   life,   the   user   should   perform   the  
following activities every six months: 

Table basis

Check screw joints and protections on pivot bolts 

Check whether the lock and table moving system work 

Check the hydraulic pump

Check whether gas springs work correctly

Table top

check whether the trendelenburg mechanism works correctly 

check whether the blockade of footrest fixing clamp works properly 

check whether gas springs work correctly

Accessories

check functionality

check whether blockades work correctly

Potential equalising 
cable

check the condition of insulation (cracks)

check how good the connection is

A checkout should involve a visual inspection and the noticed malfunctions should be handled as 
provided in 

item 6.6 Location of defects and their removal

.

 

6.6.

Location of defects and their removal

A list of repairs which may be carried out by the client on his own.

damage

Possible cause

removal

The table 
cannot be 
moved

locked wheels

Unlock blockade pedal

The 
trendelenburg 
mechanism 
does not work

The mechanism of 
trendelenburg wire 
wrongly adjusted

Regulate wire mechanism in accordance with the 
item: Table repairs and adjustment 

The table is not 
stable

Wrong adjustment 
of table feet

Unscrew or screw down feet with a wrench 

Plays in screw 
joints

Plays appeared 
during product 
operation

Screw down with generally available wrenches

_________________________________________________________________________________

page 21

Содержание SU-02

Страница 1: ...USER MANUAL Operating table SU 02 Factory no Version 12 00 ywiecka Fabryka Sprz tu Szpitalnego FAMED S A Appendixes 2 3 A List of Spare Parts...

Страница 2: ...edure was carried out in accordance with appendix VII of the directive Manufacturer ywiec Hospital Equipment Company FAMED S A Fabryczna 1 34 300 ywiec Phone numbers Infoline 48 33 866 63 75 24 hours...

Страница 3: ...er Every repair of the product must be done by a factory or an authorized service and recorded on the list of repairs which is supplied with the guarantee certificate Disregarding this requirement wil...

Страница 4: ...g the table avoid collisions Do not roll the table over electric cables If it is necessary to replace oil dangerous waste one should follow existing environmental protection regulations Notes concerni...

Страница 5: ...CHANGE OF ANGULAR POSITION OF THE FOOTRESTS 14 3 7 CHANGE OF ANGULAR POSITION OF DIVIDED FOOTRESTS 15 3 8 ROTATION OF THE FOOTREST 15 3 9 INSTALLATION AND REMOVAL OF THE FOOTREST 15 3 10 INSTALLATION...

Страница 6: ...exclusively in conformity with its destination and that it is used in appropriate conditions and in consistence with this manual The user is also obliged to take all necessary precautions in order to...

Страница 7: ...________________________________________________________________ 1 5Description of elements drawing 1 Operating table SU 02 standard version drawing 2 Operating table SU 02 Supermobile version _______...

Страница 8: ...angle change lever 17 Blockade of lever of back support angle adjustment 18 18 1 18 2 18 3 Kidney bench crank storage position adjustment position of kidney support adjustment position of lateral tilt...

Страница 9: ...m the floor 1000 mm 15 mm Backrest raise angle 70 Backrest lowering angle 40 Kidney bench raise angle 110 Footrest raise angle 4 and 5 sectional table top 15 Footrest raise angle 6 sectional table top...

Страница 10: ..._________________________________________________________________________ 1 7Use parameters The figure below indicates the most important functional parameters of the table drawing 4 _________________...

Страница 11: ...losely follow operating instruction for that equipment Improper operation may become a source of dangerous accidents There is a danger of serious burning of the patient through the contact with metal...

Страница 12: ...ng with the product manufacturer in order to ensure safe transport 2 2Unpacking and first use Caution If the product is installed by an authorised service of FAMED S A the user is released from the ob...

Страница 13: ...oes not additional operation or maintenance 3 2Longitudinal tilt of the table top to adjust longitudinal tilt of the table top trendelenburg and anti trendelenburg you should stand behind the backrest...

Страница 14: ...backrest segment Angle adjustment of the position of the back support segment Fig 1 3 position 2 is performed by muscle force with a gas spring back up In order to adjust the segment position push the...

Страница 15: ...be correctly meshed When the knob is being turned down one should check if the toothed bars are meshed 3 9Installation and removal of the footrest The footrest may be placed and removed from the seat...

Страница 16: ...he table The table in this version is equipped with central blockade of every wheel Pedal of central blockade has three positions upper direction blockade Fig 2 position 24 3 middle all wheels unblock...

Страница 17: ...ation of potential equalization in the operations room The cable is provided in standard table accessories 3 15 Collisions In some extreme positions of the table in particular when accessories install...

Страница 18: ...19 6 18 Specialist headrest WS 21 5 19 Laryngology ophthalmic headrest with track adjustment WS 21 7 20 Head rest for cervical kerb WS 21 8 21 Hand support WS 22 5 22 Body strap WS 23 0 23 gynaecologi...

Страница 19: ...e in any direction should not cause tilting or moving 2 Check mechanisms of adjustment of location of the segments controlled by gas springs by a change of position and the lock see item 3 Operation T...

Страница 20: ...roduct accessories should be reported immediately to a person in charge of such issues The bed which can not be safely operated e g damaged electric or mechanical elements must not be used till it is...

Страница 21: ...protections on pivot bolts Check whether the lock and table moving system work Check the hydraulic pump Check whether gas springs work correctly Table top check whether the trendelenburg mechanism wo...

Страница 22: ...ntification When sending asking any questions concerning the table and when ordering spare parts please give the serial number of the table placed on the nameplate and the guarantee certificate The se...

Страница 23: ...rating Table SU 02 0 _____________________________________________________________________________________________________ 7 2 Labels __________________________________________________________________...

Страница 24: ...ing and disinfection you should not use agents containing active oxygen and chlorine 5 Maximum authorized workload 150 kg 6 Attention read user manual 7 Longitudinal tilt Trendelenburg anti Trendelenb...

Страница 25: ...____________ 13 Table top height adjustment 14 Table s brake 15 Central wheel lock Notice The producer reserves the right to introduce in the product modifications resulting from technical progress wh...

Страница 26: ...C047310000000000 3 14 Potential balance cable 1 S1115000004258 2 15 Pin of potential balance 1 S11155000004256 2 16 Sliding sleeve 4 S30250100000200 2 17 Guiding sleeve 4 C073600000014000 2 18 Trende...

Страница 27: ...k rest rail I 2 C047311000020000 2 47 Crank set 1 C067204000000000 2 48 Protection 1 C067200000005000 3 49 Arm 1 C080300000700000 2 50 Screw 1 C080300000011000 2 51 Hand II 1 C047311000011000 2 52 Pro...

Страница 28: ...list Symbol SU 02 Edition 06 04 2007 Package should contain Operating table SU 02 set without headrest 1 pce and Headrest 1 pce SG 10 0 Potential equalization cord 1 pce index S11155000015625 Additio...

Отзывы: