Operating Manual for Operating Table SU-02.0
_____________________________________________________________________________________________________
-
release the central blockade of wheels (item 3.10
Table mobility
)
-
with an assistance of a few people, carefully and slowly remove the table from the pallet, then
block the wheels
-
add the headrest
(item 3.11 Installation and operation of the headrest)
-
place the product in its destination which meets the requirements described in this manual (
item
1.2 General requirements
),
-
on the basis of information given in item
3 Operation
check whether the table works as described
in item
5 Criteria on whose basis it is assessed whether product operation is correct or not
.
-
Caution!
If the product is not fully functional, i.e. the output parameters differ from the
description contained in this manual, the bed must not be used. This situation
should be reported to the producer or supplier. The use of an improperly
functioning product may result in damages, which will encumber the user and
which will not be a matter of producer’s responsibility
3 Operation and use
3.1 Raising and levelling of table top
Height of table top is changed within the scope defined in technical data by pressing of the lever
(drawing 1-3, item 10) of the hydraulic pump. Table top is lowered when the lever is raised. The foot
pump installed in the table does not additional operation or maintenance.
3.2 Longitudinal tilt of the table-top
to adjust longitudinal tilt of the table-top (trendelenburg and anti-trendelenburg), you should stand
behind the backrest, put both hands on the handles (drawing 1-3 item 13 and 14), turn the right handle
clockwise till you feel resistance (which releases the blockade of gas springs) and place table top in
the required position (within the range of angles given in technical data). When you release the right
handle, the table will be locked in the position. Additionally after aggressive discharge of the right
handle an automatic return of its rotation blockade should follow. That is belonged to make sure for
this fact through another test of handle turning (to the right) without blockade releasing, rotation of the
handle should be impossible for execution.
Neurosurgical table top
Standard table top
Fig. 6
In neurosurgical table top the handle which is placed on the right side of the back rest segment should
be put forward and turn to the headrest side.
It belongs to dismiss the handle after receiving of demanded side.
Caution!
When performing longitudinal (Trendelenburg or reverse Trendelenburg) and lateral
tilting
, patient should be secured against uncontrolled sliding down from the
operating table. Use shoulder supports, supporting rollers, side supports, belts
and grips as securing elements.
Make sure, that the rotation blockade of the right handle of longitudinal tilt
prevents its rotation!