FTP3000 Series Programmable DC Power Supply Operation Manual
____________________________________________________________________________
- 18 -
Table 3-1 Set Menu Layout
System
IP
Default 192.168.1.123
S-Mask
Default 255.255.255.0
Baud
Serial port baud rate, 4800/9600/19200/38400/115200, default: 9600.
Parity
None/odd parity/even parity, default: None
Sound
Keyboard sound ON/OFF. Default: ON
Power Save
Power down save function ON/OFF. Default: ON
Language
Simplified Chinese/Traditional Chinese/English. Default: English.
Shortcut
Fast recall function ON/OFF. Default: Off
Power On
Define power supply output On/Off state at its power on moment. Default: Off
Output
V_MAX
Upper limit of static VI output voltage set , default 100%F.S. output voltage
V_MIN
Lower limit of static VI output voltage set, default 0
I_MAX
Upper limit of static VI output current set, default 100%F.S. output current
I_MIN
Lower limit of static VI output current set, default 0
V_SLEW
Set Voltage slew rate for static VI output, range: 0.001V/s~5000V/s, default 5000V/s
I_SLEW
Set current slew rate for static VI output, range: 0.001A/s~2000A/s, default 2000A/s
Protect
OVP
Set overvoltage protection threshold, set 0 to turn off this protection, default 0
OCP
Set overcurrent protection threshold, set 0 to turn off this protection, default 0
OPP
Set overpower protection threshold, set 0 to turn off this protection, , default 0
FoldBack
Monitor output mode, protection is triggered when mode changes, settable among
choices: OFF / CC-to-CV / CV-to-CC, default set OFF
Fold-Time
Output mode changes must last for a set time for protection to be triggered. Time range:
0.1s to 600.0s
LVP
Set low voltage protection threshold, default set 0 to turn off this protection.
Reverse
Turn on/off reverse voltage protection, default value: OFF
Comm_Dog
Set communication time out protection, range: 0.0~3600.0s, set 0 to turn off this
protection;
APP Set
Prio_Slt
Choice for CV or CC priority start. Default value: Volt (CV Priority)
EXT-Single
Choose external control signal behavior, Trigger/Toggle/Hold, default: Trigger
APG
Choose external programming mode, Off/U/I/U+I, default: Off.
Response
Choose response speed in CP output mode, range 1~100%, default 50%
Sense
Choose V-sense type, local/remote, default local
Comm Pro
Choose communication protocol, SCPI/Modbus, default Modbus
DevAddr
Set device address in Modbus protocol, range 0~254
Factory
Reset to factory setting. Restart the instrument for changes to take effect.