FTP3000 Series Programmable DC Power Supply Operation Manual
____________________________________________________________________________
- 34 -
Figure 3-21 Battery Simulating Output (RESI)
3. Set the corresponding test parameters;
4. Press
to turn on battery simulating output.
Table 3-7 RESI Test Parameter Explanation
Parameter
Notes
V_MAX
RESI simulating output max voltage, range 0 ~ Vmax
I_MAX
RESI simulating output max current, range 0 ~ Imax
BAT-RES
RESI simulating battery inner resistance, range 0 ~ Vmax/Imax
3.18 RAMP Output
FTP3000 series power supply allows for output voltage ramp-up and ramp-down according to the set slew rate.
RAMP test function has three different test ways: Continuous (CONT), pulse (PULS) and toggle (TOGG).
Figure 3-22 RAMP Test
RAMP test continuous way
In continuous way, the power supply continuously switches between CV-1 and CV-2 according to the set pulse
width. Unless you exit the ramp test or turn off the power supply output, the power supply will continue to
execute according to the set parameters. This method is not affected by the trigger signal.