FTP3000 Series Programmable DC Power Supply Operation Manual
____________________________________________________________________________
- 12 -
2.4 Screen Display
Figure 2-3 MAIN PAGE Display
①
Status prompt area;
②
Readback voltage, current, etc. parameters display area;
③
Parameter setting area;
④
Multi-function key indication;
About Status prompt area
:
Output mode: CV, CC
Analog programming: Blank (analog program is Off), P-U, P-I, P-UI
V-SENSE: Blank (local sensing), Sense (Remote sensing)
Protection: Blank (no protection occurred), PFCF, SDF, OT, OCP, OPP, LVP, AMP, E01, E02, OC, OP, OV, OVP, OCVP,
RVP, COMF
Output indicator: Off (output turned off), On (output turned on)
About readback V, I parameters display area:
Displays readback voltage, current, power, measurement result etc.. However, the actual displayed information
varies, for example, in voltage/current static output mode, the area displays output voltage, current and power; in
sequence test mode, the area displays output voltage, current, sequence current steps and sequence cycle times.
About parameters setting area:
1. Rotate the rotary knob or press the direction key to move the cursor to the parameter to be edited;
2. Press “Enter” key to enter into parameter edit state;
3. Press numeric number key or rotate the rotary knob to input value;
4. Press “Enter” to validate the input and exit editing state.
About multi-function key indication:
: switch to VI (constant voltage/current) output mode, cursor is in voltage parameter;
: switch to VI (constant voltage/current) output mode, cursor is in current parameter;
: work with number key together for performing functions written underneath;
: Turn on/off power supply output.