7-4.
Cleaning and care of the product
1.
Cleaning the refrigerator filter
※
Be sure to turn OFF the power switch and the leak
and overcurrent breaker and turn OFF the master
power supply, then start operation.
Clogged filter may cause deterioration of the
refrigeration capacity. This may also cause failure
of the refrigerator.
Clogging of filter differs depending on the ambient
environment and the use time. Clean the filter
regularly according to the use conditions.
(1) Loosen the urea resin screw and remove the
ventilation cover.
(2) Take out the filter and wash it with water or neutral
detergent.
(3) After washing, dry the filter well and replace it to
the original position.
(4) Install the ventilation cover and tighten urea resin
screws.
※
Do not use the dryer because the filter may be
molten.
2.
Cleaning the product
※
Be sure to turn OFF the power switch (leak breaker)
and the master power supply before starting the
work.
Clean the main unit with a moistened and wrung off
soft cloth. For stubborn dirt, use mild detergent and
completely wipe off remaining detergent with a soft
cloth.
CAUTION
WARNING
Cooling fin
Filter
Ventilation cover
Urea resin screw
WARNING
Do not touch the cooling fin with
bare hand.
Do not touch the cooling fin with bare hand
during maintenance work. The cooling fan is
keen and may cut your hand.
Never attempt to disassembly
the product.
The unit contains parts with high voltage
applied
or
may
become
hot,
and
disassembly may cause an electrical shock
or an injury.
Use a correct method and items for
cleaning or caring the product.
When cleaning or maintaining the product,
never splash water to the exterior or the inside
directly, do not put any foreign materials and
never use the cleanser, thinner, oil, kerosene,
acid, and equivalent. Otherwise, the user may
suffer electric shock or damage to the product.
- 34 -
Содержание FDS-1000