Symptom
Cause
Countermeasure
No light exhaust sound in
about 30 seconds after start
of the vacuum pump
(
Pressure not lowering)
Vacuum cock not fully closed
Check the direction of vacuum cock and
close it fully.
Faulty installation of the trap
Retry installation of the trap.
Dust adhering to the contact surface
between the chamber cover and
packing
Remove dust
。
Light exhaust sound, but the
pressure does not lower.
Vacuum pump deteriorating in
performance because it has drawn
acid and water.
Change oil. If the trouble persists, repair
the vacuum pump.
Water and ice remaining in the trap
Remove ice, then discharge water from the
drain.
The specified amount of oil not filled
in the vacuum pump
Check the oil level and add to the specified
level.
Refrigerator switch not ON
Turn ON the refrigerator switch.
Refrigerator switch is ON, but not
enough refrigeration
Refer to the item “Faulty refrigeration.”
Pressure not returning to the
normal level even after stop
of the vacuum pump
Clogged leak filter
Replace the leak filter.
(
Refer to 7-1.Replaceing the
consumables.
)
A
la
rm
D
is
p
la
y
ed
Refrigerator alarm
issued
Self-diagnosis function
of refrigerator active
Ref.Alarm lamp ON
and ”Measured
temperature” of trap
temperature display
flashing.
・
The operating ambient
temperature has exceeded 35
℃
.
・
The system is operating with the
heat load exceeding the
refrigeration capacity.
・
Dust adhering to the air filter
・
Refrigerator fan not running
・
Low supply voltage
・
Refrigerator operated under
overload
※
When multiple causes are
involved, the operation state
differs.
・
When the ambient temperature is high,
set the room temperature to 35
℃
or
less.
・
Reduce the heat load.
・
Clean the air filter.
・
Check the refrigerator fan.
・
Check the supply voltage.
・
Before operation, set the longer
refrigerator stop period.
Temperature sensor
break alarm issued
Break in the temperature sensor
circuit
Stop operation immediately and contact the
shop from which you have purchased the
product or our nearest service center.
Lower-limit temperature
alarm issued
・
Measured temperature lowering
to “-115.0
℃
” or less
・
Ambient temperature too low
・
Long-extended no-load operation
・
Automatic recovery when the measured
temperature exceeds “-115.0
℃
” or
more.
・
When the ambient temperature is low,
set the refrigeration capacity selector
switch to “Normal.”
・
When recovery is not made, contact the
shop from which you have purchased
the product or our nearest service
center.
Alarm
Ref.Alarm
℃
Trap
Temp.
Trap temperature display
Alarm lamp
Alarm
Ref.Alarm
℃
Trap
Temp.
Trap temperature display
Alarm lamp
Alarm
Ref.Alarm
℃
Trap
Temp.
Trap temperature display
Alarm lamp
- 30 -