54
WT.050.590.000.DE.IM.0719
Functions
SFC
4.
Preparing the electrodes
1
Screw the injection cap on the bottle with the fill solution.
2
Pour a little fill solution into the filling hole. Then rinse the
electrode to wet the electrode body O-ring.
3
Press on the electrode cap until the electrode body is
loosened slightly from the epoxy cover so that the cone is also
wet.
4
Release the cap.
If the protective cover does not immediately return to its
original position, check whether the O-ring is wet enough and
repeat steps 2 and 3 until the protective cover returns.
5
The electrode must be filled with electrolyte up to the filling
hole.
6
Before use, the electrode should be placed in a 100 mg/l
fluoride solution at pH 7 for approx. 24 hours (e.g. calibration
solution).
A cap
B spring
C cable
D O-ring
E vent
F
filling hole
G cone
H reference element
I
protective cover
J
chamber for fill solution
K electrode body
Please note
The filling hole or vent must remain open. The solution in the
electrode may not be contaminated by other liquids.
Installing the fluoride sensor in
the flow block assembly
1
Install the fluoride sensor in the corresponding opening on the
flow block assembly in the Y-style flow through assembly.
2
Connect the sensor to the sensor cable marked „F
-
“.
3
The fluoride sensor can be calibrated for the first time after
approx. two to three hours running-in time.
Please note
The calibration must be checked after one day.
A
B
C
D
E
F
G
H
I
J
K
Содержание WALLACE & TIERNAN SFC
Страница 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Страница 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Страница 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Страница 121: ...SFC Installation 5 121 Top hat rail assembly...
Страница 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 123: ...SFC Installation 5 123 Top hat rail assembly...
Страница 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Страница 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Страница 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Страница 128: ...128 WT 050 590 000 DE IM 0719 Installation SFC 5 Installation of the front panel on the control cabinet...
Страница 145: ...SFC Installation 5 145 Example 2 W2T505525 W3T171924 W2T505525 Only operate with sample water inlet...
Страница 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Страница 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Страница 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Страница 249: ...SFC Complete Units Retrofit Kits and Spare Parts 8 249 Drawing of DEPOLOX 5 pressurized version W3T170032 W3T170032...
Страница 253: ...SFC Complete Units Retrofit Kits and Spare Parts 8 253 VariaSens drawing non pressurized version W3T158742 W3T158742...
Страница 257: ...SFC Complete Units Retrofit Kits and Spare Parts 8 257 VariaSens drawing pressurized version W3T158756 W3T158756...
Страница 263: ...SFC Complete Units Retrofit Kits and Spare Parts 8 263 Drawing of DEPOLOX Pool pressurized version W3T158599 W3T158599...
Страница 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Страница 273: ...SFC Declarations and certificates 10 273 10 Declarations and certificates 10 1 Declaration of Conformity...
Страница 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Страница 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...