5
Rinse off the escaped electrolyte with water.
Please note
The membrane cap (E) must be completely screwed (hand-tight)
onto the electrode shaft (B), so that no gap remains between the
two! After a run-in period of about one hour, the membrane sensor
is sufficiently run-in for an initial calibration to take place.
Calibration should be repeated after one day.
Inserting into the flow-through
adapter
1
Insert the membrane sensor through the cover into the flow-
through adapter until it is resting mechanically on the inflow
mating connector and therefore receives a good flow. It may
be necessary to turn the flow-through adapter cover to set it in
the right position relative to the inflow mating connector.
Please note
Remove air bubbles from the membrane by lifting the membrane
sensor. They interfere with the measurement!
2
Connect the measuring signal cable to the measuring device.
Please note
If the measuring sensor is installed this way, it will function reliably
for approx. three to six months.
Storing the membrane sensor
1
Lift elastomer seal (H) and only then screw off the membrane
cap (E).
2
Wash the membrane cap (E) and electrode finger (D) with
clean (distilled) water.
3
Carefully dry the electrode finger (D) with absorbent paper.
4
Leave the membrane cap (E) to dry in a dust-free place.
5
Screw the dry membrane cap (E) loosely onto the electrode
shaft (B).
Please note
The membrane (G) may not touch the gold working electrode (J).
Restarting
See „Commissioning“.
Содержание WALLACE & TIERNAN SFC
Страница 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Страница 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Страница 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Страница 121: ...SFC Installation 5 121 Top hat rail assembly...
Страница 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 123: ...SFC Installation 5 123 Top hat rail assembly...
Страница 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Страница 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Страница 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Страница 128: ...128 WT 050 590 000 DE IM 0719 Installation SFC 5 Installation of the front panel on the control cabinet...
Страница 145: ...SFC Installation 5 145 Example 2 W2T505525 W3T171924 W2T505525 Only operate with sample water inlet...
Страница 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Страница 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Страница 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Страница 249: ...SFC Complete Units Retrofit Kits and Spare Parts 8 249 Drawing of DEPOLOX 5 pressurized version W3T170032 W3T170032...
Страница 253: ...SFC Complete Units Retrofit Kits and Spare Parts 8 253 VariaSens drawing non pressurized version W3T158742 W3T158742...
Страница 257: ...SFC Complete Units Retrofit Kits and Spare Parts 8 257 VariaSens drawing pressurized version W3T158756 W3T158756...
Страница 263: ...SFC Complete Units Retrofit Kits and Spare Parts 8 263 Drawing of DEPOLOX Pool pressurized version W3T158599 W3T158599...
Страница 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Страница 273: ...SFC Declarations and certificates 10 273 10 Declarations and certificates 10 1 Declaration of Conformity...
Страница 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Страница 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...