216
WT.050.590.000.DE.IM.0719
Operation
SFC
6.
Position. Ym?
Ym range too narrow
Position of positioning motor is
calibrated wrong
Check the distance between the
calibration points.
Positioner selected, but not
connected
Check setting: Positioner with Ym
Feedback signal incorrect
Check DIP switch for feedback
Positioner feedback incorrectly
connected or defective
Check (refer to 9. “Wiring Diagrams”)
Module?
Sensor module has been removed
Sensor module defective
Refit or replace the sensor module
Adaption?
Adaption terminated with error
Refer to 4.9 “Adaption”.
Range?
Min. or max. limit value is outside
the measuring range
Check the min/max limit values and
change, if necessary
mA-Input 1 ?
mA-Input 2 ?
mA-Input signal has been exceeded
or fallen short of
Check mA connection or signal
Ym calibration?
Positioner calibration incorrect
Check calibration of the positioner
CAN
measurement?
No CAN bus station present
Check CAN bus, configure CAN station
CAN-actuator?
No CAN bus actuator present
Check CAN bus, configure CAN station
Ym display blinking
Positioner disconnected
Switch positioner to automatic
Sample Line Dos
Automatic sample water inlet
disinfection
Time-controlled function is ended
automatically, as soon as the process is
complete
Error message
Cause
Remedy
Содержание WALLACE & TIERNAN SFC
Страница 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Страница 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Страница 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Страница 121: ...SFC Installation 5 121 Top hat rail assembly...
Страница 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 123: ...SFC Installation 5 123 Top hat rail assembly...
Страница 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Страница 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Страница 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Страница 128: ...128 WT 050 590 000 DE IM 0719 Installation SFC 5 Installation of the front panel on the control cabinet...
Страница 145: ...SFC Installation 5 145 Example 2 W2T505525 W3T171924 W2T505525 Only operate with sample water inlet...
Страница 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Страница 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Страница 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Страница 249: ...SFC Complete Units Retrofit Kits and Spare Parts 8 249 Drawing of DEPOLOX 5 pressurized version W3T170032 W3T170032...
Страница 253: ...SFC Complete Units Retrofit Kits and Spare Parts 8 253 VariaSens drawing non pressurized version W3T158742 W3T158742...
Страница 257: ...SFC Complete Units Retrofit Kits and Spare Parts 8 257 VariaSens drawing pressurized version W3T158756 W3T158756...
Страница 263: ...SFC Complete Units Retrofit Kits and Spare Parts 8 263 Drawing of DEPOLOX Pool pressurized version W3T158599 W3T158599...
Страница 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Страница 273: ...SFC Declarations and certificates 10 273 10 Declarations and certificates 10 1 Declaration of Conformity...
Страница 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Страница 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...