4.
4.2.1
3-electrode sensor DEPOLOX
®
5 and DEPOLOX
®
Pool
Flow block assembly
DEPOLOX
®
5 3-electrode
measurement for free Cl
2
, ClO
2
,
O
3
or KMnO
4
Potable, industrial and swimming pool water are disinfected
almost exclusively by adding chlorine, chlorine dioxide, ozone or
potassium permanganate.
With the flow block assembly DEPOLOX
®
5 with integrated 3-
electrode cell, the contents of this disinfectant can be continuously
recorded.
A sensor module ("DES" for 3-electrode cells) and terminal strips
are used to connect the flow block assembly DEPOLOX
®
5 to the
SFC. Various controller functions are available depending on the
application selected. The flow block assembly DEPOLOX
®
5 is
also used for holding other sensors, such as pH, Redox, fluoride,
conductivity or membrane sensors for free chlorine, chlorine
dioxide, ozone, total chlorine, or combined chlorine.
A pressurized and non-pressurized version with flow rate control is
available as a flow block assembly with integrated 3-electrode
cells (see "Overall Function" page 31).
Mode of operation of the 3-
electrode sensor in the flow
block assembly DEPOLOX
®
5
The measuring cell in the flow block assembly DEPOLOX
®
5 is a
3-electrode cell with external potentiostatic control circuit. Working
and counter electrodes are designed as half-ring electrodes and
consist of a special platinum alloy.
The reference electrode is a silver/silver chloride electrode, which
is connected to the sample water via two diaphragms
(membranes). The reference electrode with PVC support is
immersed into an electrolyte solution.
The electrolyte supply can be replenished during operation if
necessary (see7.2 “Maintenance of flow block assembly
DEPOLOX® 5”).
By connecting the 3-electrode cell to the SFC (DES sensor module
for 3-electrode cell), a variable Upot cell voltage can be output
between the working electrode (red) and reference electrode
(white) via the potentiostatic control circuit. A measuring cell
current (µA signal), which is evaluated using the SFC, sets itself
proportional to the disinfectant concentration in the sample water.
A special cleaning sand is filled into the flow block assembly, which
is circulated by the sample water current and continuously cleans
the platinum electrodes.
A multi-sensor is integrated into the flow block assembly
DEPOLOX
®
5 to measure the temperature and monitor the flow
rate. This is made of a stainless steel housing and is used
simultaneously as the sample water grounding (for connection to
SFC, see 9. “Wiring Diagrams”).
Содержание WALLACE & TIERNAN SFC
Страница 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Страница 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Страница 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Страница 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Страница 121: ...SFC Installation 5 121 Top hat rail assembly...
Страница 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 123: ...SFC Installation 5 123 Top hat rail assembly...
Страница 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Страница 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Страница 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Страница 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Страница 128: ...128 WT 050 590 000 DE IM 0719 Installation SFC 5 Installation of the front panel on the control cabinet...
Страница 145: ...SFC Installation 5 145 Example 2 W2T505525 W3T171924 W2T505525 Only operate with sample water inlet...
Страница 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Страница 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Страница 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Страница 249: ...SFC Complete Units Retrofit Kits and Spare Parts 8 249 Drawing of DEPOLOX 5 pressurized version W3T170032 W3T170032...
Страница 253: ...SFC Complete Units Retrofit Kits and Spare Parts 8 253 VariaSens drawing non pressurized version W3T158742 W3T158742...
Страница 257: ...SFC Complete Units Retrofit Kits and Spare Parts 8 257 VariaSens drawing pressurized version W3T158756 W3T158756...
Страница 263: ...SFC Complete Units Retrofit Kits and Spare Parts 8 263 Drawing of DEPOLOX Pool pressurized version W3T158599 W3T158599...
Страница 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Страница 273: ...SFC Declarations and certificates 10 273 10 Declarations and certificates 10 1 Declaration of Conformity...
Страница 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Страница 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Страница 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...