W3T331759 / WT.050.585.003.UE.IM.1014
56
MICRO/2000® AND DEOX/2000® MODULE
WARNING:
Hydrochloric acid and muriatic acid will severely irritate
eyes and cause chemical burns to skin. Always wear appropriate protective
equipment during handling. When diluting acid, always pour acid into the
water, never pour water into acid. It is the responsibility of the user of the
equipment to obtain and follow the safety precautions of the manufacturer
of the hazardous material.
If the probe is to be stored for more than a week, the porous element should
be removed, the electrolyte drained, and the reservoir should be rinsed
with distilled water to eliminate any remaining electrolyte. Otherwise the
electrode end must be kept wetted to prevent crystallization of the electrolyte
in the pores of the element and clogging of the element.
The porous element can be removed and the electrolyte drained as follows:
Using a self-tapping screw (a #6 drywall screw is preferred), tap into the
porous element, then pull out the element. The element is approximately
6mm (¼-inch) long.This porous element is no longer usable. The electrolyte
will drain out slowly.Rinse the reservoir well with distilled water to remove
residual electrolyte. When removing the element in this manner, use care
not to scar the sides of the element bore as this might prevent insertion of a
new piece of porous material or cause excessive leakage of the electrolyte.
A new porous element can be created and installed as follows: Do not
precut the porous material to length. A clean, sharp razor blade is required
to make a clean cut in the porous material. First, using a sawing motion,
slice off a thin slice of one end of a length of the porous material. Press this
fresh cut end into the element bore. The fit should be snug. The material
should be embedded to a 6 to 9mm (¼" to 3/8") depth. Cut off protruding
portion of porous element flush with probe tip, being very careful not to
scratch or damage electrodes.
Содержание WALLACE & TIERNAN DEOX/2000
Страница 4: ......
Страница 10: ...W3T331759 WT 050 585 003 UE IM 1014 6 MICRO 2000 AND DEOX 2000 MODULE...
Страница 30: ...W3T331759 WT 050 585 003 UE IM 1014 26 MICRO 2000 AND DEOX 2000 MODULE...
Страница 46: ...W3T331759 WT 050 585 003 UE IM 1014 42 MICRO 2000 AND DEOX 2000 MODULE...
Страница 74: ...W3T331759 WT 050 585 003 UE IM 1014 70 MICRO 2000 AND DEOX 2000 MODULE...