background image

Important Notice

Evoqua Water Technologies Instruction Manuals are the safety documents for their products. The information provided therein 

is in accordance with the best practices known at the time of issue and, if fully implemented, should enable the user to operate 

the plant in the safest possible manner. It is therefore most important that instructions be read and understood prior to the 

installation, operation or maintenance of the plant.

Notes

Where necessary, special instructions are provided and are 

quoted under the following headings:

 WARNING:

 A warning is given in instances where 

failure to observe the instruction could result in injury to 

personnel.

 CAUTION:

 Cautions are given where failure to 

observe the instruction may result in equipment damage, 

or pollution to an allied system.

NOTE: Notes are provided to give additional emphasis to 

particular points of detail.

The following recommendations are made to ensure safe 

and efficient operation:-

1  Only suitably qualified personnel should install, operate 

and maintain the equipment.

2  Only Evoqua Water Technologies manufactured or 

approved parts should be used.

3  The equipment should not be used for any purpose other 

than that for which it was supplied. 

4  If equipment being supplied is being put into storage or 

not being commissioned immediately, or if the plant is being 

de-commissioned, Evoqua Water Technologies urge that 

they be contacted for detailed advice.

In an effort to make progressive improvements to products, 

changes in design may be incorporated from time to time, which 

may not be reflected immediately in the instruction manual. 

If in doubt, contact Evoqua Water Technologies quoting the 

equipment serial number; serial numbers are essential for 

effective communication and proper equipment identification.

The equipment described in this manual may be used with 

substances that themselves may be hazardous to personnel 

safety. It is essential that persons employed in the vacinity of 

such substances be aware of the appropriate safety practices 

and the location of suitable safety equipment. Provision of 

safety cards and codes of practice concerning hazardous 

substance should be made by the supplier of the substance.

Warranty

The equipment supplied is guaranteed against mechanical 

defect notified to the Seller within a period of 1 year from 

the date of delivery. For such guarantee to be valid the 

purchaser must notify the seller in writing immediately 

such defect becomes apparent. If the Seller so requires, 

the Purchaser shall return the defective equipment (at the 

Purchaser’s expense) to the Seller. The Seller undertakes 

that it will, at its own option and its own expense, and by way 

of full discharge of its guarantee obligation hereunder, either 

repair, or supply a replacement for the defective equipment, 

or refund any purchase monies paid to it in respect of any 

such defective equipment, or portion thereof. If title to the 

defective equipment is at the time of such replacement or 

refund vested in the Purchaser, then title shall thereupon 

vest in the Seller. Replacement material situated outside 

the U.K. will be supplied F.O.B. U.K. Port. The guarantee 

given in this clause shall not be operative and enforceable 

if the equipment is not operated strictly in accordance 

with the sellers instruction, or in the respect of any defect 

arising from accident, deliberate act, misuse, neglect or 

from a breach from the terms of the following clause, or 

in respect of any defect arising through damage incurred 

while being transported after delivery of the equipment, or if 

the equipment has been altered or modified in any way by 

any person other than the seller, or in respect of any other 

cause whatsoever which lies beyond the Sellers control.

The Purchaser must comply with all user instructions 

and safety recommendations issued by the Seller, and 

must install, commission and maintain the equipment in 

accordance with good engineering practice, and under the 

supervision of suitably qualified personnel, and the Seller 

shall not be liable to the Purchaser for any loss suffered 

as a result of the Purchaser’s breach of the terms of this 

clause.

Subject to this guarantee all conditions, warranties and 

representations, whether expressed or implied (by statute 

or otherwise) relating to the equipment, are hereby excluded 

in so far as they can be excluded without such exclusion 

being void or unenforceable at law.

Wallace & Tiernan, Barrier, Chem-Ad, DEPOLOX, OSEC, and V10k are trademarks of Evoqua, its subsidiaries or affiliates. 

The designations used in this publication may be trademarks whose use by third parties for their own purposes could 

violate the rights of the owners.All rights, especially those to duplication and distribution as well as translation, are 

reserved. No part of this document may be reproduced in any form (printing, photocopying, microfilm or any other 

method) or saved, processed, duplicated or distributed by the use of electronic systems without the express written 

consent of Evoqua Water Technologies.

Содержание WALLACE & TIERNAN DEOX/2000

Страница 1: ...nan Instruction Manual MICRO 2000 and DEOX 2000 Measurement Module W3T331759 WT 050 585 003 UE IM 1014 Suite 3 2nd Floor One 60 London Road Sevenoaks TN13 1BT Tel 44 1732 771777 Fax 44 1732 771800 www...

Страница 2: ...must notify the seller in writing immediately such defect becomes apparent If the Seller so requires the Purchaser shall return the defective equipment at the Purchaser s expense to the Seller The Se...

Страница 3: ...il Australia 61 0 3 8720 6597 info au evoqua com Canada 1 905 944 2800 canadainfo evoqua com Germany 49 8221 9040 wtger evoqua com Singapore 65 8630 7100 infosg evoqua com U S A 1 856 507 9000 wtus ev...

Страница 4: ......

Страница 5: ...y or secondary wastewater or cooling water In a DEOX 2000 system it operates as a centre zero analyser capable of measuring oxidants Chlorine and Ozone and reductants Sulphur Dioxide Potassium Iodide...

Страница 6: ...der and under the operating conditions indicated in the technical specifications Compliance with the intended use also includes reading this operating manual and observing all the instructions it cont...

Страница 7: ...zard to such persons or third parties and damage to the system or unit itself or to other property are possible Work described in this operating manual may only be performed by authorized personnel Pe...

Страница 8: ...gents and replaced part Warranty Conditions Thefollowingmustbeobservedforcompliancewithwarrantyconditions Installation commissioning by trained and authorized personnel Intended use Observation of the...

Страница 9: ...ing any changes in the operating performance In the event of malfunctions switch the unit off immediately Have malfunctions remedied immediately Installation and Maintenance Work Always perform instal...

Страница 10: ...W3T331759 WT 050 585 003 UE IM 1014 6 MICRO 2000 AND DEOX 2000 MODULE...

Страница 11: ...nsor input 3 electrode cell Principle of operation Potentiostatic amperometry Temperature drift 0 2 Linearity error 0 1 Cell voltage 250 mV Cell current 1 to 7 A mg l Temperature input PT 1000 Measuri...

Страница 12: ...ture 16 to 70 C 4 to 158 F 1 1 1 MICRO 2000 and DEOX 2000 flow block assembly Housing Facade Dimensions W x H x D mm 330 x 330 x 254 Weight approx 9 1 kg Connections Sample water 6mm OD hose Thread co...

Страница 13: ...ure 10 to 30 C 14 to 86 F Max pressure 0 4 bar 5 psi at pump suction Water quality clean water potable water quality municipal industrial waste seawater Flow 7 6l h 2USg h asconstantaspossible into sa...

Страница 14: ...nding on the application they can be operated either without a flow block assembly no sensor measuring module or in conjunction with a flow block assembly and sensor measuring module Flow Block Assemb...

Страница 15: ...MICRO 2000 and DEOX 2000 flow block assembly The MICRO 2000 and DEOX 2000 flow block assembly contains the following Cell body with cover Flow control valve mounted externally By pass strainer mounte...

Страница 16: ...585 003 UE IM 1014 12 MICRO 2000 AND DEOX 2000 MODULE Figure 1 3 A Cover B Cell body with sensor C Peristaltic pump D Inlet shut off valve E Bypass Y strainer with flow control valve F Inlet control v...

Страница 17: ...strations Dimensions Top Hat Rail Assembly MICRO 2000 and DEOX 2000 Flow Block 50 590 100 050 Wall Mount Assembly MICRO 2000 and DEOX 2000 Flow Block 50 590 100 060 SFC Schematic Wiring 50 585 155 050...

Страница 18: ...the mounting rail to the wall with two screws 2 Hook the flow block assembly onto the mounting rail to the left of the electronic module and fasten to the wall with two screws CAUTION The flow block...

Страница 19: ...w see Figure 2 3 e Replace probe in sample cell NOTE All the grit in the sample cell should be in suspension any not in suspension should be removed Figure 2 1 Figure 2 2 Figure 2 3 2 3 3 Insert the s...

Страница 20: ...etering pump while power is applied to the pump The motor has enough torque to cause severe pain and injury if fingers are caught in the pumping mechanism WARNING To avoid possible severe personal inj...

Страница 21: ...ng Connections Cut the excess reagent tubing and connect to the reagent fittings at the inlet to the reagent tube units Install the T fitting provided in the tubing at the inlet to bottom of the analy...

Страница 22: ...applications the sample may be run through 6mm inch pipe or tubing and the bypass flow on the Y strainer may be shut It should be opened only intermittently to flush the strainer element and sample l...

Страница 23: ...e end of the drain fitting and run hose to a waste drain b Connect a hose from theY strainer bypass valve to the waste drain c Insidethewetside inserttheanalyserbypasstubingandcelldischarge tubing bot...

Страница 24: ...make contact with one another Never connect or disconnect any cables to which voltage is applied NOTE Refer to manual W3T168777 for SFC fusing details and manual W3T159207 for the MFC fusing details...

Страница 25: ...21 W3T331759 WT 050 585 003 UE IM 1014 MICRO 2000 AND DEOX 2000 MODULE TOP HAT RAIL ASSEMBLY DIMENSIONS MICRO 2000 and DEOX 2000 Flow Block 50 590 100 050 ISSUE 1 7 14...

Страница 26: ...W3T331759 WT 050 585 003 UE IM 1014 22 MICRO 2000 AND DEOX 2000 MODULE WALLMOUNT ASSEMBLY DIMENSIONS MICRO 2000 and DEOX 2000 Flow Block 50 590 100 060 ISSUE 1 7 14...

Страница 27: ...r board SFC Alarm 1 4 Power L2 N L1 1 Shield earthed at one end Temp Digital In Feedback signal Analog in Ym DI1 DI2 control stop 0 4 20mA 0 4 20mA 0 4 20mA control constant 0 1V mA input 1 mA input 2...

Страница 28: ...VAC L1 L2 220 240 VAC prefuse max 230 115V AC 6A Relay Outputs MICRO 2000 DEOX 2000 Wet Side Power Connection 24 23 22 21 20 19 MFC SCHEMATIC WIRING 50 585 155 050B ISSUE 1 1 11 MICRO DEOX 2000 WETSID...

Страница 29: ...2000 AND DEOX 2000 MODULE Main Board 3 pot cells temp sensor 1 2 3 4 5 6 7 8 3 4 WRK REF CNT WH shield BK RD GRN BRN Pt1000 MICRO 2000 DEOX 2000 1 Shield earthed at one end SFC Sensor Module Connectio...

Страница 30: ...W3T331759 WT 050 585 003 UE IM 1014 26 MICRO 2000 AND DEOX 2000 MODULE...

Страница 31: ...le water industrial process water and waste water Typical applications Measurement and monitoring of water parameters Chlorine single feedback closed loop control Possible process measurements are fre...

Страница 32: ...MICRO 2000 and DEOX 2000 flow block electrodes Optimum flow around all sensors 3 2 Measurement Inputs In principle the following sensor measuring module types or retrofit kits can be installed at the...

Страница 33: ...H2 O Cl I2 The analyser measuring process is as follows A continuous sample is delivered to the analyser A flushing Y strainer dividesthesampleintotwostreams thelargerbypassstreamcontinuously flushes...

Страница 34: ...etering a predetermined amount of Potassium Iodate along with pH4 buffer and potassium iodide into a metered quantity of process water sample This constant level addition of iodine is termed the Iodin...

Страница 35: ...erred to as the reaction tubing The reacted reagents are then mixed with the sample inlet to the metering pump and the Biased sample then enters the analyser cell A three electrode probe is immersed i...

Страница 36: ...elements or the porous element white element as finger dirt or oil may impede the probe operation The platinum material is thin and easily punctured or sheared off The electrodes are not repairable Al...

Страница 37: ...Calibration 4 4 1 MICRO 2000 DEOX 2000 Calibration 4 4 2 Errors 4 5 4 1 Display and Operator Controls Graphic display and operating panel All information is shown on the graphic display Analyser 1 DI...

Страница 38: ...sor menu Move forwards or backwards by six hours in the trend graph Escape Cancel the entry without saving the new value Move up one menu level Enter Acknowledge Acknowledge alarm message Set the runn...

Страница 39: ...the display with the up or down arrow keys Confirm the entry with ENTER Cancel entry Press the ESC key to exit the menu item Entries which have not been confirmed are reset to their original settings...

Страница 40: ...asic display in the main menu select the Calibration menu 2 Select the Temperature menu item The Calib temperature window appears on the graphic display 3 Select Cal temperature 4 Perform comparative...

Страница 41: ...emand free water 3 Allow the demand free water to pump through the sample line for a minimum of 10 minutes using the 3 way inlet valve see Figure 4 3 4 Once the reading is stable enter this value as Z...

Страница 42: ...inal Bias value is 1 times the range of the analyser Due to variations in feed rate the actual Bias can vary 25 a Place 500 ml of demand free water in a 500 ml flask and place the flask in the wet sid...

Страница 43: ...the unit with higher range units having a higher signal strength Once the signal is stable enter the desired bias value 1 x range Return to main display and allow unit to operate on demand free water...

Страница 44: ...e cells Sensor has zero current 5 A or 5 A Electrodes in the 3 electrode cell are soiled clean and service if necessary Sample water is turned off or sample line leaks Upot potential voltage set incor...

Страница 45: ...inlet disinfection Time controlled function is ended automatically as soon as the process is complete Error The following table shows and explains possible errors which can occur If you are unable to...

Страница 46: ...W3T331759 WT 050 585 003 UE IM 1014 42 MICRO 2000 AND DEOX 2000 MODULE...

Страница 47: ...nd replacement 5 2 9 Probe electrolyte and electrodes 5 2 10 Probe cleaning refilling and porous element replacement 5 2 11 5 1 Maintenance Schedules Thefollowingmaintenanceschedulesarerecommendations...

Страница 48: ...calibrate each time the cell sand is replaced The calibration must be checked after one day 5 2 Maintaining MICRO 2000 and DEOX 2000 Flow Block Assembly WARNING To avoid electrical shock turn off powe...

Страница 49: ...theanalyserbeforeremovingtheocclusionringfromthe metering pump The occlusion ring stops the free flow of liquid through the analyser sample lines If it is removed with flow still present at the analys...

Страница 50: ...Replacement 1 Place the occlusion ring loosely in place over the tube units but do not attempt to clamp the occlusion ring in place yet 2 Hold the ring in place with one hand and slide a bolt with wa...

Страница 51: ...ack face of the body and loosen the set screw one or two turns NOTE Hold the panel upright while removing the rotor body to avoid dropping and losing the rollers To use as a guide for reinstallation t...

Страница 52: ...the analyser sample lines If it is removed with flow still present at the analyser process water will be sprayed from the pump discharge line and from the cell top Sprayed process water can create an...

Страница 53: ...are replaced with the tube units c Removetheocclusionringmountingbolts requiresa7 16 inchsocket d Remove the tube units Installation of tube units refer to Figure 5 1 Figure 5 1 1 Remove the occlusion...

Страница 54: ...should be against the outside surfaces of tube retainer 6 Checkthatthetubeunits BorC areintheirpropersmall H orlarge I groove in the roller G Check that the reagent tubes do not cross 7 Replacetheoccl...

Страница 55: ...sea3 32 inchAllenwrenchtoloosenthesetscrewholdingtheimpeller on the shaft This set screw leaves a noticeable mark on the shaft If the shaft is not replaced reinstall the impeller in the same position...

Страница 56: ...lide the grit guard small end first onto the shaft and up against the seal adapter 7 Slide the impeller onto the shaft up to the grit guard Retighten the set screw until just snug 8 Install the cell b...

Страница 57: ...the back of the panel before reinstalling the panel Motor replacement is the reverse sequence of removal 5 2 8 Impeller shaft replacement 5 2 8 1 Impeller Shaft Removal 1 Disconnect power to the wetsi...

Страница 58: ...ntity of silicone grease to the O ring and impeller shaft 8 Reinstall the seal adapter grit guard cell sealing O ring impeller and cell body and cell retaining screws as described in paragraph 5 2 6 I...

Страница 59: ...f excessive use of electrolyte becomes a nuisance requires constant monitoring of the level replace the porous element and check the element bore for damage that would hinder the sealing around the el...

Страница 60: ...ainedasfollows Using a self tapping screw a 6 drywall screw is preferred tap into the porous element then pull out the element The element is approximately 6mm inch long Thisporouselementisnolongerusa...

Страница 61: ...List of Contents Section MICRO 2000 and DEOX 2000 CE Wetside Enclosure Asssembly 50 580 000 035A B MICRO 2000 and DEOX 2000 Wetside Casting Assembly 50 580 000 040A B Micro CE DEOX 2000 wetside GAA11...

Страница 62: ...W3T331759 WT 050 585 003 UE IM 1014 58 MICRO 2000 AND DEOX 2000 MODULE MICRO 2000 AND DEOX 2000 CE WETSIDE ENCLOSURE ASSEMBLY PARTS W3T140748 240V W3T332446 115V 50 580 000 035A ISSUE 2 7 14...

Страница 63: ...16882 1 Mach d mounting plate 13 W2T365269 4 Mounting plate screw 14 W2T365269 1 10 32 Unc x 3 8 green ground screw 15 W2T366020 2 Mb3a10c2 cable mounts 16 W2T19230 2 Terminal block mounting screw 17...

Страница 64: ...W3T331759 WT 050 585 003 UE IM 1014 60 MICRO 2000 AND DEOX 2000 MODULE MICRO 2000 AND DEOX 2000 WETSIDE CASTING ASSEMBLY PARTS W3T140748 240V W3T332446 115V 50 580 000 040B ISSUE 3 7 14...

Страница 65: ...21 2 Screw 10 24 x 3 4 lg 19 W2T13299 3 Clip plastic 20 W2T12538 1 Bearing pvc 21 W2T12112 1 U cup teflon 22 W2T418119 1 Seal grit teflon 23 W2T11682 1 Coupling brass 24 W2T11674 1 Occlusion ring tefl...

Страница 66: ...arrangements to reduce water flow during testing Connect cold water supply to the analyser inlet Water supply should be set to 0 2 bar 3 PSI higher pressure may result in water spray in unit Connect r...

Страница 67: ...E N D I 0 2 0 G N I B U T 1 P60424 W2T11670 TEE REDUCTION CONNECTOR KYNAR PVDF 1 ITEM PT NUM SAP NUM DESCRIPTION MATERIAL QTY W3T108408 AAD3506 2 REAGENT PARTS GROUP 6 P60420 W2T11668 1 F D V P R A N...

Страница 68: ...2 4 W2T14737 2 Elbow pvc 90 5 W2T12055 2 Union nut pvc 1 2 20 thread 6 W2T16217 1 Tubing polyethylene 1 4 id 3 ft 7 W2T15468 1 Valve cci pvc 8 W2T13310 1 Y strainer pvc 9 W2T12996 1 Labcock valve pvc...

Страница 69: ...65 W3T331759 WT 050 585 003 UE IM 1014 MICRO 2000 AND DEOX 2000 MODULE W3T108351 WATERLINE KIT PARTS LIST 50 580 000 050B ISSUE 1 9 10...

Страница 70: ...34 1 Rigid tubing 027 ID 13 lg 5 W2T16958 2 1 4 lockwasher SS 6 W2T18825 2 Mach Screw rd Hd SS 1 4 20 x 3 4 lg 7 W2T417965 2 Hex jam nut SS 1 4 20 8 W2T14861 1 Strap 9 W2T13382 1 Bottom bracket shelf...

Страница 71: ...050 585 003 UE IM 1014 MICRO 2000 AND DEOX 2000 MODULE 7 COMPLETE DEVICES RETRO FIT KITS SPARE PARTS List of Contents PARA NO Retrofit Sets 7 1 Spare Parts and Consumables 7 2 MICRO 2000 and DEOX 2000...

Страница 72: ...maintenance kit W3T107926 s1 Sample tube assembly Sample metering pump W3T107927 s2 Reagent tube assembly Sample metering pump U10242 s1 Tube silicone grease for impeller shaft seal W2T12538 s1 Beari...

Страница 73: ...cetic acid U86584 W2T636701 5 litres de ionised water U86586 W2T636702 Potassium iodide 250g AAA8160 W2T633474 Potassium iodate solution AAB1128 W2T633562 MICRO 2000 Old Part No New Part No Chemical p...

Страница 74: ...W3T331759 WT 050 585 003 UE IM 1014 70 MICRO 2000 AND DEOX 2000 MODULE...

Страница 75: ...oncentration The analyser measuring process is as follows A continuous sample is delivered to the analyser A flushing Y strainer dividesthesampleintotwostreams thelargerbypassstreamcontinuously flushe...

Страница 76: ...skin or clothing Wash thoroughly after handling In case of spill soak up with sand or earth When servicing liquid reagent lines pumps or refilling reservoirs turn off pumps to avoid possible squirtin...

Страница 77: ...f the analyser The quantity of KI required is calculated as follows 50 4 range grams KI Example Analyser range 5 mg l 50 4 5 70 grams KI CAUTION Never mix potassium iodide KI with sulphuric acid as fr...

Страница 78: ...servoir Use a funnel when filling reagent reservoirs The use of a funnel allows the reservoir to be filled without removing the rigid length of tubing from the reservoir If the tubing is removed from...

Страница 79: ...ignificant bromide residual e g in saline waters Bromine residual behaves very similarly to chlorine residual and so the application recommendations for each are the same 4 1 3 Chlorine residual monit...

Страница 80: ...h organic chlorine compounds in the measurement of potassium permanganateorchlorinedioxideresiduals ammoniumsulphatereagent liquidisusedtoeliminatetheresponse interference tofreechlorinewhen necessary...

Страница 81: ...HSO3 A range of chlorine species is typically present in industrial processes theserangingfromhypochlorousacid HOCl tocombinedchlorinesuch aschloramines ThereactionwithSO2 forthesespeciesisindicatedbe...

Страница 82: ...especially if there is a long process time An alternative approach is to use a centre zero analyser such as the SFC DEOX 2000 In this approach an iodine bias is applied to the cell Iodine I2 is reduce...

Страница 83: ...any area of contact immediately If chemicals contact eyes rinse with copious amounts of water and seek immediate medical attention WARNING pH4bufferandaceticacidwillseverelyirritateeyesandcause chemi...

Страница 84: ...assium Iodide KI solution is prepared by dissolving 166g KI in 5 litres of distilled water The Potassium Iodate KI03 reagent is prepared by adding KI03 concentrate to the Acetic acid The table below i...

Страница 85: ...nology for water treatment in the municipal industrial and the leisure industries AFTERMARKET SUPPORT AND SERVICE Meeting our customers needs is our highest priority In addition to supplying the produ...

Страница 86: ...och EastDunbartonshire E Belfast NorthernIreland F Limerick Ireland TRAINING Evoqua Water Technologies understands the importance of training to our customers receiving the correct training can ensure...

Отзывы: