W3T331759 / WT.050.585.003.UE.IM.1014
14
MICRO/2000® AND DEOX/2000® MODULE
2.1.2 Storage
Store the unit and the sensors in a dry condition without any residual water
in a dry place. Storage temperature, see section 1.1, "Specifications".
2.2 Installation
The device must be protected against rain, frost and direct sunlight and
should not be installed outdoors. It must be mounted horizontally on a
flat wall with an ambient temperature of 0 to 50°C (32 to 122°F). The
air in the room should be non-condensing.
2.2.1 Opening the housing
1. Remove the housing cover of the flow block assembly.
2.2.2 Installation with mounting rail (see drawing 50.590.100.050)
1
Fasten the mounting rail to the wall with two screws.
2.
Hook the flow block assembly onto the mounting rail to the left of
the electronic module and fasten to the wall with two screws.
CAUTION:
The flow block assembly does not need to be mounted
directly next to the electronics, it can be mounted on separate mounting
rail. The exact location limited by available probe cable lengths.
2.2.3 Installation without mounting rails
When wall mounting the unit it should be screwed to a flat surface through
the holes in the corners of the casing, the hole centres for the mounting
being 254 mm x 254 mm.
2.3 Commissioning
2.3.1 Installation guide
Commissioning procedure:
Refer to manual W3T168777 for SFC and manual W3T159207 for the
MFC.
2.3.2 Pour in the cell sand
Adding grit to MICRO/2000® and DEOX/2000® flow block.
a. Remove probe from sample cell.
b. Wet finger tip.
Содержание WALLACE & TIERNAN DEOX/2000
Страница 4: ......
Страница 10: ...W3T331759 WT 050 585 003 UE IM 1014 6 MICRO 2000 AND DEOX 2000 MODULE...
Страница 30: ...W3T331759 WT 050 585 003 UE IM 1014 26 MICRO 2000 AND DEOX 2000 MODULE...
Страница 46: ...W3T331759 WT 050 585 003 UE IM 1014 42 MICRO 2000 AND DEOX 2000 MODULE...
Страница 74: ...W3T331759 WT 050 585 003 UE IM 1014 70 MICRO 2000 AND DEOX 2000 MODULE...