41
W3T331759 / WT.050.585.003.UE.IM.1014
MICRO/2000® AND DEOX/2000® MODULE
Cell ?
In 3 electrode cells:
Chlorine sensor not
connected properly
No sand cleaning
Sensor, sensor cable or
sensor module defective
Sensor measuring
module μA measuring
range exceeded
Connect sensor correctly.
Check sand cleaning; add if
necessary
Check the sensor, sensor
cable or sensor module,
replace if necessary
Select higher μA measuring
range
Range?
Min/max limit value is
outside
the measuring range
Check the min/max limit
values and change, if
necessary
Sample Line Dos Automatic sample water
inlet disinfection
Time-controlled function is
ended automatically, as soon
as the process is complete
Error
The following table shows and explains possible errors which can occur.
If you are unable to remedy the error yourself, please contact the Evoqua
Water Technologies service department.
Table 4.2 - Errors
Error
Cause
Remedy
No indication on
device
No power supply
External switch or fuse off
Device fuse defective
Check the power supply and
replace fuse
Housing cover is fitted
incorrectly (MFC units
only)
Check, fit the housing cover
correctly (cable possibly
trapped)
Measured value
display not
available, although
the appropriate
measuring
module is installed
Measuring module
defective or fitted
incorrectly
Check, refit module correctly,
replace measuring module
Содержание WALLACE & TIERNAN DEOX/2000
Страница 4: ......
Страница 10: ...W3T331759 WT 050 585 003 UE IM 1014 6 MICRO 2000 AND DEOX 2000 MODULE...
Страница 30: ...W3T331759 WT 050 585 003 UE IM 1014 26 MICRO 2000 AND DEOX 2000 MODULE...
Страница 46: ...W3T331759 WT 050 585 003 UE IM 1014 42 MICRO 2000 AND DEOX 2000 MODULE...
Страница 74: ...W3T331759 WT 050 585 003 UE IM 1014 70 MICRO 2000 AND DEOX 2000 MODULE...