4
ENGLISH
STERILE (ETO) unless package has been
opened or damaged.
DO NOT REUSE.
Federal (US) law restricts this device to
sale by or on the order of a physician.
DESCRIPTION
Note: VivaSight-SL is the brand name for
the ETView Tracheoscopic Ventilation
Tube (TVT
TM
).
VivaSight-SL is a single-use, cuffed PVC
endotracheal tube with an embedded
video imaging device and light source
at its tip, and an integrated single use
video/power cable (adapter cable) with
connector. VivaSight-SL displays images
of the airway onto the Ambu® aView
TM
monitor for as long as the device remains
in place during intubation.
VivaSight-SL has two Murphy eyes flanking
the 45° bevel tip. In addition to the cuffed
inflation tube, VivaSight-SL has an injection
port for imaging lens cleaning that leads to
two lumens running along the tube’s wall
and opens distally to the imaging lens.
The port is used for cleaning the lens with
air, saline, or other appropriate solutions.
INDICATIONS FOR USE
The ETView Tracheoscopic Ventilation
Tube (TVT
TM
) is intended for intubation
procedures. The ETView Tracheoscopic
Ventilation Tube (TVT
TM
) is indicated for
use as a temporary artificial airway in
adults requiring mechanical ventilation.
It is intended for oral and nasal intubations.
The TVT
TM
System is indicated for
viewing during non-difficult and difficult
intubation procedures, for verifying
endotracheal tube and endobronchial
blocker placement and repositioning,
for viewing during suctioning and for
general inspection of the airway.
CONTRAINDICATIONS
Do not use laser equipment in the
immediate proximity of the VivaSight-SL.
ACCESSORIES
Stylet
WARNINGS
•
VivaSight-SL is a single use product.
Do not reuse: After use, handle and
dispose according to local regulations
for disposal of infected medical devices.
Reuse of this single use product may
create a potential risk to the user.
Reprocessing, cleaning, disinfection
and sterilization may compromise
product characteristics which in turn
create an additional risk of physical
harm to or infection of the patient.
•
Do not attempt to resterilize any part
of VivaSight-SL.
•
Do not connect VivaSight-SL into
any device or outlet other than those
supplied by ETView or Ambu.
•
The intended performance of the
VivaSight-SL is reached only with the
disposable devices and accessories
provided or specified by Ambu or
ETView.
•
VivaSight-SL is available in tube sizes of
7 mm, 7.5 mm and 8 mm (internal
diameter) and should be used for
patients who require one of these sizes.
•
Do not use the injection port for
suctioning the patient.
•
The cuff pressure should not exceed
25 cm H
2
O. Diffusion of a nitrous oxide
mixture, oxygen, or air may either
increase or decrease the cuff pressure.
•
Do not use the camera in the event
that inspired gas temperature exceeds
34 °C for an artificially ventilated patient.
•
Do not overinflate the cuff. Overinflation
can result in tracheal damage, rupture
of the cuff with subsequent deflation,
or cuff distortion, which may lead to
airway blockage.
•
Do not use the tube if the cuff is
damaged. Care must be taken to avoid
damaging the cuff during intubation.
•
Do not use the tube if the cuff is
damaged. Various bony anatomical
structures such as teeth, or any intubation
aid with sharp surfaces in the intubation
VIVASIGHT- EB
VIVASIGHT-SL