Wave Analyzer Medica 700 User Guide
204
Appendices
Immunity
test
Test level IEC
60601
Compliance
level
Electromagnetic environment
– guidance
Electrical
fast
transient/
burst
IEC 61000-
4-4
± 2 kV for
power supply
lines
± 1 kV for input/
output lines
Complies
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-
4-5
± 1 kV
differential mode
±2 kV common
mode
Complies
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage
dips, short
interruption
s and
voltage
variations
on power
supply input
lines
IEC 61000-
4-11
<5 % UT (>95 %
dip in UT) for 0,5
cycle
40 % UT (60%
dip in UT) for 5
cycles
70 % UT (30 %
dip in UT) for 25
cycles
<5 % UT (>95 %
dip in UT) for 5
sec
Complies
Mains power quality should that of
a typical commercial or hospital
environment. If the user of the
Wave Analyzer Medica 700
requires continued operation
during power mains interruptions,
it is recommended that the Wave
Analyzer Medica 700 be powered
from an uninterruptible power
supply or a battery.
Power
frequency
(50/60
hertz)
magnetic
field
IEC 61000-
4-8
3 A/m
Complies
Power magnetic fields should be
at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note:
UT is the a.c. mains voltage prior to application of the test level.
Содержание Medica 700
Страница 1: ...Version 8 2 2018 28 Wave Analyzer Medica 700 2018 Visionix User Guide UM WAM 700 MAY 2018 Rev2 2...
Страница 2: ......
Страница 7: ...1 Introduction I...
Страница 10: ...Wave Analyzer Medica 700 User Guide 4 Introduction Copyright 2018 Manufacturer All rights reserved...
Страница 11: ...2 Safety II...
Страница 15: ...3 Equipment and Installation III...
Страница 25: ...4 Overview of the Software IV...
Страница 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Страница 94: ...5 Managing Patients V...
Страница 103: ...6 Performing a Diagnostic VI...
Страница 128: ...7 Managing the List of Test Results VII...
Страница 135: ...8 Managing the Database VIII...
Страница 141: ...9 General Actions and Features IX...
Страница 147: ...10 Configuring the Unit X...
Страница 189: ...11 What Should I Do If XI...
Страница 191: ...12 Maintenance XII...
Страница 194: ...Wave Analyzer Medica 700 User Guide 188 Maintenance 2 Click on Turn Off button 3 Click on OK button...
Страница 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Страница 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Страница 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Страница 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Страница 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Страница 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Страница 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Страница 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Страница 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Страница 204: ...13 Appendices XIII...