IEC60601-1 ed.3.1.:
Medical electrical equipment - Part 1: General requirements for
basic safety and essential electrical safety performance.
IEC60601-1-2 ed.3.:
Medical electrical equipment - Part 1-2: General requirements for
safety - Collateral standard: Electromagnetic compatibility – Requirements and tests.
Lifetime of the product: 7 years
1rst CE marking : 2013-12
Electromagnetic Emissions
The Wave Analyzer Medica 700 is intended for use in the electromagnetic environment
specified below. The customer or the user of the Wave Analyzer Medica 700 should assure
that it is used in such an environment.
Emission test
Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Wave Analyzer Medica 700 uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Wave Analyzer Medica 700 is suitable for use
in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purpose.
Harmonic
emissions CEI
61000-3-2
Class A
Voltage
fluctuations /
flicker emissions
CEI 61000-3-3
Complies
Electromagnetic Immunity
The Wave Analyzer Medica 700 is intended for use in the electromagnetic environment
specified below. The customer or the user of the Wave Analyzer Medica 700 should assure
that it is used in such an environment.
Immunity
test
Test level IEC
60601
Compliance
level
Electromagnetic environment
– guidance
Electrostati
c discharge
(ESD)
IEC 61000-
4-2
±6 kV contact
± 8 kV air
Complies
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Содержание Medica 700
Страница 1: ...Version 8 2 2018 28 Wave Analyzer Medica 700 2018 Visionix User Guide UM WAM 700 MAY 2018 Rev2 2...
Страница 2: ......
Страница 7: ...1 Introduction I...
Страница 10: ...Wave Analyzer Medica 700 User Guide 4 Introduction Copyright 2018 Manufacturer All rights reserved...
Страница 11: ...2 Safety II...
Страница 15: ...3 Equipment and Installation III...
Страница 25: ...4 Overview of the Software IV...
Страница 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Страница 94: ...5 Managing Patients V...
Страница 103: ...6 Performing a Diagnostic VI...
Страница 128: ...7 Managing the List of Test Results VII...
Страница 135: ...8 Managing the Database VIII...
Страница 141: ...9 General Actions and Features IX...
Страница 147: ...10 Configuring the Unit X...
Страница 189: ...11 What Should I Do If XI...
Страница 191: ...12 Maintenance XII...
Страница 194: ...Wave Analyzer Medica 700 User Guide 188 Maintenance 2 Click on Turn Off button 3 Click on OK button...
Страница 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Страница 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Страница 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Страница 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Страница 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Страница 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Страница 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Страница 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Страница 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Страница 204: ...13 Appendices XIII...