Wave Analyzer Medica 700 User Guide
46
Overview of the Software
Special Alerts
The summary tab will display alerts when results fall outside the expected range:
No
Description
1
Refraction
Refraction is displayed in red:
When there is a difference of more than 2D in the refraction between the eyes,
or the cylinder difference between the eyes is more than 2D.
If the difference in refraction of the same eye in mesopic and photopic light is a
different of 1D sphere and 0.75D cylinder, the caution is also displayed.
When Sphere is above
±
20D or Cylinder is more than
±
8D.
When pupil diameter is less than 2.5mm in at least one eye.
When Pupil size difference between OD and OS is greater than 1mm.
When cylinder is greater than 4D in either eye.
2
Add
Add is displayed in red when the difference between the expected addition for age
and measured addition is great than 1D.
3
E Simulation
E simulation is outlined in red when:
The delta between the night and day sphere is >1D and/or cylinder is greater
than >0.75D
The Ocular HOA RMS equivalent sphere is larger than 0.5D in the right or left
eye.
Содержание Medica 700
Страница 1: ...Version 8 2 2018 28 Wave Analyzer Medica 700 2018 Visionix User Guide UM WAM 700 MAY 2018 Rev2 2...
Страница 2: ......
Страница 7: ...1 Introduction I...
Страница 10: ...Wave Analyzer Medica 700 User Guide 4 Introduction Copyright 2018 Manufacturer All rights reserved...
Страница 11: ...2 Safety II...
Страница 15: ...3 Equipment and Installation III...
Страница 25: ...4 Overview of the Software IV...
Страница 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Страница 94: ...5 Managing Patients V...
Страница 103: ...6 Performing a Diagnostic VI...
Страница 128: ...7 Managing the List of Test Results VII...
Страница 135: ...8 Managing the Database VIII...
Страница 141: ...9 General Actions and Features IX...
Страница 147: ...10 Configuring the Unit X...
Страница 189: ...11 What Should I Do If XI...
Страница 191: ...12 Maintenance XII...
Страница 194: ...Wave Analyzer Medica 700 User Guide 188 Maintenance 2 Click on Turn Off button 3 Click on OK button...
Страница 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Страница 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Страница 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Страница 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Страница 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Страница 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Страница 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Страница 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Страница 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Страница 204: ...13 Appendices XIII...