201
Wave Analyzer Medica 700 User Guide
201
Appendices
Tonometer Specifications
Calibrated Range
7mmHg – 44mmHg
Accuracy
+/-2 mmHg
Method
Non-contact air puff applanation tonometry
Retroillumination Specifications
Measurement wavelength
850nm
Method
Retro led
LCD Screen Specifications
Screen type
TFT LCD
Touchscreen type
Capacitance type
Size
10” WSVGA resolution (1024 × 600 pixels)
Contrast ratio
500
Typical luminosity
250 Cd/m2
In a clinical comparison with a Goldmann tonometer on 231 eyes covering the range 7 –
33 mmHg, 95% of measurements with the Wave Analyzer Medica 700 tonometer were
within ±5mmHg of the Goldmann tonometer measurements.
13. 2 Clinical Studies
Pachymeter Study:
A clinical study was performed including 68 normal eyes that compared the Wave Analyzer
Medica 700 pachymeter function with the Pentacam rotating Scheimpflug camera.
Measurements of central corneal thickness (CCT) were recorded. The mean CCT on the
Wave Analyzer Medica 700 was 541.74 ±39.76µm (std dev) and on the Pentacam was 535.72
± 37.93µm. The Wave Analyzer Medica 700 measures 6.02µm thicker than the Pentacam. A
Bland Altmann graph showing the average and standard deviations of the measurement
differences as well as the 95% confidence levels (std dev × 1.96) and the 99% confidence
levels (std dev × 2.58) is shown in Figure A1. The 95% confidence levels for the measurement
were from 24.54µm to -16.80µm and 99% confidence levels for the measurement were from
Содержание Medica 700
Страница 1: ...Version 8 2 2018 28 Wave Analyzer Medica 700 2018 Visionix User Guide UM WAM 700 MAY 2018 Rev2 2...
Страница 2: ......
Страница 7: ...1 Introduction I...
Страница 10: ...Wave Analyzer Medica 700 User Guide 4 Introduction Copyright 2018 Manufacturer All rights reserved...
Страница 11: ...2 Safety II...
Страница 15: ...3 Equipment and Installation III...
Страница 25: ...4 Overview of the Software IV...
Страница 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Страница 94: ...5 Managing Patients V...
Страница 103: ...6 Performing a Diagnostic VI...
Страница 128: ...7 Managing the List of Test Results VII...
Страница 135: ...8 Managing the Database VIII...
Страница 141: ...9 General Actions and Features IX...
Страница 147: ...10 Configuring the Unit X...
Страница 189: ...11 What Should I Do If XI...
Страница 191: ...12 Maintenance XII...
Страница 194: ...Wave Analyzer Medica 700 User Guide 188 Maintenance 2 Click on Turn Off button 3 Click on OK button...
Страница 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Страница 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Страница 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Страница 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Страница 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Страница 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Страница 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Страница 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Страница 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Страница 204: ...13 Appendices XIII...