Wave Analyzer Medica 700 User Guide
184
What Should I Do If...?
You may encounter some of the following common problems while working with the Wave
Analyzer Medica 700. If you do, try following the suggested solutions listed below. If the problem
persists and the suggested solution does not remedy it, contact a qualified service
representative or your local distributor.
Problem
Possible Causes
Suggested Solutions
Images are not
properly centered
External lights or
windows near the
unit
Large patches of
light near the unit
Ensure that there no bright lights or
windows near the unit or the patient.
Ensure that there are no large
irregular patches of light on the patient
or on the unit.
Measured pupil size
is different from the
expected or actual
pupil size
External lights or
windows near the
unit
Large patches of
light near the unit
Ensure that there no bright lights or
windows near the unit or the patient.
Ensure that there are no large
irregular patches of light on the patient
or on the unit.
High-order
aberrations that
differ from what is
expected
Direct light falling
on the unit or on
the patient.
Ensure that there no bright lights or
windows near the unit or the patient.
Ensure that there are no large
irregular patches of light on the patient
or on the unit.
The Pachy
measurement
always goes to
manual mode
The pachy filter is
foggy due to the
cold
Use the anti-fog spray at least once a
week, and when the pachy
consistently goes to manual mode.
Tickets do not print
Paper jam
Out of paper
Open the print paper door and release
the paper.
Replace the paper roll.
Содержание Medica 700
Страница 1: ...Version 8 2 2018 28 Wave Analyzer Medica 700 2018 Visionix User Guide UM WAM 700 MAY 2018 Rev2 2...
Страница 2: ......
Страница 7: ...1 Introduction I...
Страница 10: ...Wave Analyzer Medica 700 User Guide 4 Introduction Copyright 2018 Manufacturer All rights reserved...
Страница 11: ...2 Safety II...
Страница 15: ...3 Equipment and Installation III...
Страница 25: ...4 Overview of the Software IV...
Страница 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Страница 94: ...5 Managing Patients V...
Страница 103: ...6 Performing a Diagnostic VI...
Страница 128: ...7 Managing the List of Test Results VII...
Страница 135: ...8 Managing the Database VIII...
Страница 141: ...9 General Actions and Features IX...
Страница 147: ...10 Configuring the Unit X...
Страница 189: ...11 What Should I Do If XI...
Страница 191: ...12 Maintenance XII...
Страница 194: ...Wave Analyzer Medica 700 User Guide 188 Maintenance 2 Click on Turn Off button 3 Click on OK button...
Страница 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Страница 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Страница 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Страница 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Страница 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Страница 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Страница 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Страница 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Страница 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Страница 204: ...13 Appendices XIII...