Wave Analyzer Medica 700 User Guide
106
Performing a Diagnostic
Troubleshooting
If your Camera View looks
like this
What to do
The cross is not focused, the cross is not visible, only
one white circle is visible. Press on the
continuously until the cross becomes focused. If the
cross becomes blurry again, you passed the focus,
press on the
button until it becomes clear again.
When if focus, the tono will fire the puff automatically.
The circle is centered, the patient moved. Push the the
house icon
to get back to your initial point and
then continue to focus and center the cross and circle.
The cross is not focused, press the
continuously
until the cross becomes better focused. If the cross
becomes blurry again, you passed the focus, press the
button until satisfied with the focus. The tono will
fire automatically at the correct focus.
Focused and centered but no measurement taken?
If the circles and cross are focused and clear, but the device has not taken a measurement,
you can force the device to take a measurement by pressing Force Fire
.
If the device remains in one location, and does not seem to be autofocusing, check
to see if the device is set to
Semi
mode.
Semi Mode
By default, the device has an Auto focus which is fast and accurate. However, another focusing
mode available is Semi Mode. In this mode, the tonometer will remain at the focal point found
by the pachymeter, and correction of the focus must be done by the examiner. To adjust the
focus, follow the steps listed above. To select the tonometer modes, go to
.
Please note, in this mode, the device will remain at a set focus point, and all the
focusing will be done by the examiner, no tono measurements will be taken if the
device is not in focus
.
Demo Puff/Modify number of measurements
If you would like to demonstrate the tono puff for your patient, select the Parameters button and
then press "Demo Puff". This popup menu also allows you to modify the number of times each
eye is measured by the tonometer.
149
149
Содержание Medica 700
Страница 1: ...Version 8 2 2018 28 Wave Analyzer Medica 700 2018 Visionix User Guide UM WAM 700 MAY 2018 Rev2 2...
Страница 2: ......
Страница 7: ...1 Introduction I...
Страница 10: ...Wave Analyzer Medica 700 User Guide 4 Introduction Copyright 2018 Manufacturer All rights reserved...
Страница 11: ...2 Safety II...
Страница 15: ...3 Equipment and Installation III...
Страница 25: ...4 Overview of the Software IV...
Страница 38: ...Wave Analyzer Medica 700 User Guide 32 Overview of the Software Ringer Screen...
Страница 94: ...5 Managing Patients V...
Страница 103: ...6 Performing a Diagnostic VI...
Страница 128: ...7 Managing the List of Test Results VII...
Страница 135: ...8 Managing the Database VIII...
Страница 141: ...9 General Actions and Features IX...
Страница 147: ...10 Configuring the Unit X...
Страница 189: ...11 What Should I Do If XI...
Страница 191: ...12 Maintenance XII...
Страница 194: ...Wave Analyzer Medica 700 User Guide 188 Maintenance 2 Click on Turn Off button 3 Click on OK button...
Страница 195: ...189 Wave Analyzer Medica 700 User Guide 189 Maintenance 4 Head and ChinRest are lowered 5...
Страница 196: ...Wave Analyzer Medica 700 User Guide 190 Maintenance 6 7...
Страница 197: ...191 Wave Analyzer Medica 700 User Guide 191 Maintenance 8 9...
Страница 198: ...Wave Analyzer Medica 700 User Guide 192 Maintenance 10 11...
Страница 199: ...193 Wave Analyzer Medica 700 User Guide 193 Maintenance 12...
Страница 200: ...Wave Analyzer Medica 700 User Guide 194 Maintenance 13...
Страница 201: ...195 Wave Analyzer Medica 700 User Guide 195 Maintenance 14...
Страница 202: ...Wave Analyzer Medica 700 User Guide 196 Maintenance 15...
Страница 203: ...197 Wave Analyzer Medica 700 User Guide 197 Maintenance 16...
Страница 204: ...13 Appendices XIII...