Operation
StamoLys CA 71FE
22
Hauser
Analog output 1
0 ... 20 mA / 4 ... 20 mA
4 ... 20 mA
Selection for the scope of channel 1 measuring
range. If the concentration measuring range is
0 ... 5 mg/l, this corresponds to 0 mg/l either 4 mA
or 0 mA. The end of the measuring range is the
same in both cases at 20 mA.
Analog output 2
Two-channel version only!
Selecting scope of
channel 2 measuring range.
The scopes of the measuring range are
independent of each other for channel 1 and
channel 2 and are determined by the start of
measuring range (channel 1 / channel 2) or end of
measuring range (channel 1 / channel 2) setting in
the PARAMETER ENTRY menu.
Alarm value AV 1-1
NO current
NC current
!
Note!
Changes will only be activated
after a Reset (Power off/on)!
Setting for whether contact for alarm value 1,
channel 1 works as NO current or NC current
contact.
Alarm value AV 2-1
Setting for whether contact for alarm value 2,
channel 1 works as NO current or NC current
contact.
Alarm value AV 1-2
Two-channel version only!
Setting for whether contact for alarm value 1,
channel 2 works as NO current or NC current
contact.
Alarm value AV 2-2
Two-channel version only!
Setting for whether contact for alarm value 2,
channel 2 works as NO current or NC current
contact.
Error contact
Setting for whether error contact works as
NO current or NC current contact
Current date/time
01.01.96 00:00... 31.12.95 23:59
Setting the system clock.
Format DD.MM.YY hh:mm.
Calibrate offset
yes / no
Frequency offset
Pressing the
J
key starts a blind value
measurement for compensating the reagent’s
inherent colour.
Frequency offset
– 5000 ... +5000
0
Manually changing the frequency offset.
Option
Range of adjustment
(default settings in bold)
Display
Info
Analog output 1
4-20 mA
Analog output 2
4-20 mA
Alarm val. 1-1
norm. closed
Alarm val. 2-1
norm. closed
Alarm val. 1-2
norm. closed
Alarm val. 2-2
norm. closed
Error contact
norm. closed
act. Date/Time
25.01.02 15:45
Calibrate offs
yes: K
no: E
f-Offset
[Hz]
0