StamoLys CA 71FE
Commissioning
Hauser
29
The following summary presents procedures carried out in the device and the intervals
you have programmed:
6.2.2
Wet commissioning
This is different to the dry commissioning in that, for wet commissioning the reagent
lines are filled before the automatic measurement, calibration and flushing cycle is
started.
Proceed as follows:
1.
Plug-in the plug into a socket.
2.
Press the
H
key until AUTO MEASURING is displayed.
3.
Select the SERVICE menu.
4.
Switch the P2 reagent pump "on" (with
F
select P2 and with
V
set to"g") and leave
it running until you can tell that there are reagents at the T-connector. After this,
switch P2 "off" (s) again with
W
.
5.
Now switch the valves to passage for standard (select V1: S, V2: S; with
F
or
K
switch to "S" with
V
) and then switch the P1 sample pump "on". Leave the pump
running until you can tell that there is standard at the T-connector. Switch P1 back
"off".
6.
Now switch the valves to passage for cleaning agent (select V1: S, V2: R; with
F
or
K
switch to "R" or "S" with
V
) and then switch the P1 sample pump "on". Leave
the pump running until you can tell that there is cleaning agent at the T-connector.
Switch P1 back "off".
7.
Now switch the valves to passage for sample (select V1: P, V2: S; with
F
or
K
switch to "P" or "S" with
V
) and then switch the P1 sample pump "on". Leave the
pump running until you can tell that there is sample at the T-connector. Switch P1
back "off".
!
Note!
For the two-channel version, valve V3 must be set additionally for switch-over between
channel 1 and channel 2.
8.
Now proceed as with dry commissioning (from step 2).
Function
Duration [s]
Range of adjustment
Measurement
Flushing (sample)
Delay to sample
Stabilisation
1st measurement
Flushing (reag. line)
Fill mixture
Reaction
2nd measurement
Flushing (sample)
3 x 15
20 ... 300
4
2
15 ... 18
s. Techn. data
30
CONFIGURATION / "Delay to sample"
SERVICE / "Mixture"
Calibration
Flushing (standard)
Delay to standard
Stabilisation
1st measurement
Flushing (reag. line)
Fill mixture
Reaction
2nd measurement
Flushing (sample)
3 x 15
20 ... 300
4
2
15 ... 18
s. Techn. data
30
CONFIGURATION / "Delay to sample"
SERVICE / "Mixture"
Flushing
Pump cleaning solution
Allow to react
Pump cleaning solution
Flush hold on: 2
5
Flush hold on: 2
PARAMETRIEREN / "Flush hold on"