StamoLys CA 71FE
Wiring
Hauser
17
Two-channel version
!
Note!
Condition fulfilled means:
•
AV 1: concentration > Alarm value 1
•
AV 2: concentration > Alarm value 2
•
Fault: error occurred
Contacts AV 1, AV 2 and fault are only affected during automatic operation.
4.4
Serial interface
Terminal assignment:
The results (measured value+unit of CR) are output in the "Data memory
Measured values" menu.
The calibration results (measured value+unit of CR) are output in the "Data
memory-Calibration factors" menu.
!
Note!
•
A null modem cable is required (not a crossed one).
•
The analyzer does not have to be configured for the interface.
The following commands can be sent from the PC, in order to read out data:
•
"D" = Data memory-Measured values
•
"C" = Data memory-Calibration factors
•
"S" = Setup (parameter entry, configuration...)
•
"F" = Frequency (current)
Connection
Terminal connection for
condition fulfilled
Terminal connection for
condition not fulfilled
Terminal connection for
power off
AV 1 - 1
A:
R:
25
25
-
-
23
24
A:
R:
25
25
-
-
24
23
25
-
24
AV 1 - 2
A:
R:
13
13
-
-
11
12
A:
R:
13
13
-
-
12
11
13
-
12
AV 2 - 1
A:
R:
28
28
-
-
26
27
A:
R:
28
28
-
-
27
26
28
-
27
AV 2 - 2
A:
R:
16
16
-
-
14
15
A:
R:
16
16
-
-
15
14
16
-
15
Fault
A:
R:
19
19
-
-
17
18
A:
R:
19
19
-
-
18
17
19
-
18
Channel ½
measurement end
A:
R:
22
22
-
-
20
21
A:
R:
22
22
-
-
21
20
22
-
21
A = NO current configured
R = NC current configured
RS 232 of CA 71
COM 1/2 at PC
SUB-D, nine-pin
Function
Function
SUB-D, nine-pin
3
TxD
RxD
2
2
RxD
TxD
3
8
CTS
RTS
7
CTS
8
5
GND
GND
5
Software protocol:
9600, N, 8, 1
Output format:
ASCII