Wiring
StamoLys CA 71FE
16
Hauser
4.2.3
Two-channel version
4.3
Switching contacts
One-channel version
Connection
Designation
Function
Signal inputs
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started,
display flashes
Signal outputs
AV 1 - 1
Alarm value 1, channel 1 exceeded or undershot
AV 1 - 2
Alarm value 2, channel 1 exceeded or undershot
AV 2-1
Alarm value 1, channel 2 exceeded or undershot
AV 2 - 2
Alarm value 2, channel 2 exceeded or undershot
Fault
Retrieves error message using operation menu
Channel ½ or
measurement end
1
1) Alternative
selection
Displays active channel
Displays "Measurement finished" (5 s)
Analog output
l-1 channel 1
0 or 4 mA = measuring range start
20 mA = measuring range end
l-2 channel 2
0 or 4 mA = measuring range start
20 mA = measuring range end
Channel selection
Meas. 2
0 V = channel 1
24 V = channel 2
Connection
Terminal connection for
condition fulfilled
Terminal connection for
condition not fulfilled
Terminal connection for
power off
AV 1
A:
R:
25
25
-
-
23
24
A:
R:
25
25
-
-
24
23
25
-
24
AV 2
A:
R:
28
28
-
-
26
27
A:
R:
28
28
-
-
27
26
28
-
27
Fault
A:
R:
19
19
-
-
17
18
A:
R:
19
19
-
-
18
17
19
-
18
Unassigned
22
16
13
-
-
-
20
14
11
22
16
13
-
-
-
21
15
12
22
16
13
-
-
-
21
15
12