StamoLys CA 71FE
Wiring
Hauser
15
4.2
Signal connection
4.2.1
Screening of the analog outputs
The interference suppressor attenuates electromagnetic effects on control, power and
signal lines.
After the connection of the data transfer cables clip the interference suppressor (in
scope of delivery) on the cable cores (not on the outer insulation of the cable!). Place
the cable screen out of the interference suppressor and connect it to PE (brass screw
with bolt, top right in the connection compartment) (
!
Note!
With the two-channel version, place the cable cores of all cables (data cables to analog
output 1 and to analog output 2) through the interference suppressor.
4.2.2
One-channel version
C07-CA71xxx-03-08-00-xx-001.eps
Fig. 12:
Interference protection of the signal cable
1
Cable screen (to PE )
2
Signal cable
3
Interference suppressor
4
Cable cores of the signal cable
Connection
Designation
Function
Signal inputs
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started,
display flashes
Signal outputs
AV 1
Alarm value 1 exceeded or undershot
AV 2
Alarm value 2 exceeded or undershot
Fault
Retrieves error message using operation menu
Analog output
l-1 channel 1
0 or 4 mA = measuring range start
20 mA = measuring range end