Commissioning
StamoLys CA 71FE
28
Hauser
6
Commissioning
6.1
Function check
#
Warning!
•
Check that all connections have been made correctly. Check, in particular, that all
hose connections are secure, so that no leaks occur.
•
Ensure that the mains voltage corresponds to the voltage specified on the nameplate.
6.2
Switch-on
6.2.1
Dry commissioning
!
Note!
•
If possible, let the analyzer warm up in standby mode before commissioning ("Auto
measuring" display). The time can be defined via the "1st measurement" option in the
PARAMETER ENTRY menu.
•
At the start of measurement with a cold analyzer, the first measuring results will be
errored. The reaction is temperature-dependent and if the temperature is too low the
pre-defined reaction time is insufficient for a complete reaction. For this reason, never
carry out calibration with a cold analyzer. Wait at least two hours before carrying out
calibration.
When the analyzer has been configured and calibrated, the measuring cycle starts
automatically. Entering parameters is no longer necessary.
To perform initial start-up or to readjust the device parameters, proceed as follows:
1.
Plug-in the plug into a socket.
2.
Press the
H
key until AUTO MEASURING is displayed.
3.
Select the CONFIGURATION menu and program the individual options up to and
including the "Current date/time". With
H
you can return to the main menu.
4.
Now complete the PARAMETER ENTRY and SERVICE menus.
With
H
you can return to the main menu.
5.
Select CONFIGURATION again and use
F
to go to the "Calibrate offset" option.
6.
Connect a vessel containing distilled water to the "Sample" connection and start the
frequency offset (
J
key). The registered value is displayed and saved.
7.
Then reconnect the sampling line.
With
H
you can return to the main menu.
The analyzer starts the "Calibration", "Measurement", and "Flushing" procedures
automatically (triggered by control signal or integrated timer) in accordance with the
device parameters that you have set (1st calibration, 1st measurement, 1st flushing
times and the respective intervals control the temporal procedure).