Wiring
StamoLys CA 71FE
14
Hauser
4.1.2
Terminal assignment
!
Note!
•
Alarm values 1 and 2 do not need to be connected if the PLC sets its own alarm values
at the analog output.
•
When using a sample conditioning system:
Connect terminals 57 and 53 on the CA 71 to the corresponding terminals on the
sample conditioning system. For the allocation of these terminals, please see the
sample conditioning system operating instructions.
•
If there is a 24 V voltage at terminal 57, the analyzer will not begin measurement
(sample not ready). To start measurement, the voltage must stay at 0 V for at least
5 seconds.
Function
Designation
Terminal without
cooling device
Terminal with
cooling device
Mains
L
3
3
N
2
2
PE
1
1
Alarm value 1, channel 1
COM
25
25
NC
24
24
NO
23
23
Alarm value 2, channel 1
COM
28
28
NC
27
27
NO
26
26
Alarm value 1, channel 2
COM
–
13
NC
–
12
NO
–
11
Alarm value 2, channel 2
COM
–
16
NC
–
15
NO
–
14
Fault
COM
19
19
NC
18
18
NO
17
17
Reserve (unassigned terminals)
COM
22
22
NC
21
21
NO
20
20
Analog output 1
0/4 ... 20 mA
+
36
36
–
35
35
Screen
PE
1
PE
1
Analog output 2
0/4 ... 20 mA
+
–
39
–
–
38
Screen
–
PE
1
1)
Brass screw with bolt top right in the connection compartment (marked with
*
)
Sample conditioning
remote control
Input
57
57
0 V
53
53
Channel switch-over
Input
–
55
0 V
–
53