StamoLys CA 71FE
Operation
Hauser
27
5.4.2
Calibration example
Proceed as follows if you want to activate an immediate calibration (e.g. after you have
replaced the reagents).
Ensure that you have changed the reagents, filled the hoses again (no air bubbles) and
the analyzer is in measuring mode.
1.
Hold
H
down until AUTO MEASURING appears.
2.
Use
W
to move through the PARAMETER ENTRY menu and press
F
.
3.
Use
F
to go to the "1st calibration" option.
4.
Select the option with
F
5.
Now use the
W
or
V
and
F
keys to set a time which lies in the past.
6.
Press
F
to accept the value and then press
H
twice to return to the main menu
7.
Press
F
again. This takes you back to measuring mode.
The calibration is now carried out automatically.
"
Caution!
After the calibration has finished, the analyzer automatically goes into measuring mode.
You now have to set the time of the 1st calibration back into the future in order to align
the calibration and rinsing times to each other. The rinse must be performed 3-4 hours
before the next calibration.
Proceed as described above to change the setting for the time of the 1st calibration.
After changing to measuring mode, the analyzer automatically begins measuring,
rinsing and calibrating at the defined times.
Measuring range
Concentration [mg/l]
Frequency [Hz]
Iron, lower measuring range
FE-A
10 ... 500 µg/l
0.000
0.050
0.100
0.150
0.200
0.250
0.300
0.350
0.400
0.500
0
19
44
68
92
116
139
163
186
232
Iron, medium measuring range
FE-B
0.05 ... 2.00 mg/l
0.00
0.50
1.00
1.50
2.00
0
238
492
754
1010
Iron, upper measuring range
FE-C
0.10 ... 5.00 mg/l
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
4.00
5.00
0
238
492
754
1010
1251
1477
1693
1920
2307
Содержание StamoLys CA 71 FE
Страница 2: ......
Страница 51: ...StamoLys CA 71FE Appendix Endress Hauser 51...
Страница 52: ...Appendix StamoLys CA 71FE 52 Endress Hauser...