Trouble-shooting
StamoLys CA 71FE
38
Hauser
9.4
Spare parts
Fig. 15 shows the components of the analyzer. Please, take the spare parts order
numbers from the following sections.
"Measurement"
display flashes
Time of 1st
measurement not
reached
The date must be between 01.01.1996 and the current date.
Interval not expired
Change parameters.
Calibration does not
start
Time of 1st calibration
not reached
The date must be between 01.01.1996 and the current date.
Interval not expired or
0 h
Change parameters.
Leak on photometer
E+H-Service
Flushing does not
start
Time of 1st flushing
not reached
The date must be between 01.01.1996 and the current date.
Interval not expired or
0 h
Change parameters.
Leak on
Photometer
Device or discharge
blocked
Remove blockage.
E+H-Service
Blockage, deposits in
device
Water hardness
Lime deposits can by flushing with 5% hydrochloric acid .
If necessary, dose EDTA in sample stream to prevent sedimentation.
Not enough sample
conditioning
Shorten cleaning intervals of sample conditioning..
Error message
Possible cause
Tests and / or corrective measures
C07-CA71XXx-00-08-xx-006.eps
Fig. 15:
Overview parts
A
B
C
P
Sample input two-channel version
Standards and reagent canisters
Collecting vessel
Sample
R1
S
Reagent 1 canister