StamoLys CA 71FE
Technical data
Hauser
41
10.4
Performance characteristics
10.5
Environment
10.6
Process
10.7
Mechanical construction
Response time t
100
Time between two measurements:
reaction time + rinse time + waiting time + rinse again time +
filling time (minimum waiting time = 0 min)
Maximum measured error
2 % of measuring range end
Measuring interval
2 ... 120 minutes
Reaction time
2 minutes
Sample requirement
15 ml / measurement
Reagent requirement
1 x 0.15 ml / measurement
0.65 l of reagent per month with 10 minute measuring interval
Calibration interval
0 ... 72 h
Rinse interval
0 ... 72 h
Rinse time
selectable from 20 to 300 s (standard = 80 s)
Rinse again time
30 s
Filling time
22 s
Maintenance interval
6 month (typically)
Servicing requirements
15 minutes / week (typically)
Ambient temperature
5 ... 40 °C (41... 104 °F), avoid markable fluctuations
Humidity
below the condensation limit, installation in usual, clean rooms
outdoor installation only possible with protective devices
(customer supplies)
Ingress protection
IP 43
Sample flow rate
min. 5 ml per min
Consistance of the sample
low solid content (< 50 mg/l)
Sample inlet
pressureless
Design, dimensions
see chapter "Installation"
Weight
GFK housing
approx. 28 kg (61.7 lb)
Stainless steel
housing
approx. 33 kg (72.8 lb)
Materials
Housing
Stainless steel 1.4301 (AISI 304) or GFK
Front windows
Plexiglass
®
Endless hose
C-Flex
®
, Norprene
®
Pump hose
Tygon
®
, Viton
®
Valves
Tygon
®