Smartec CLD18
Technical data
Hauser
41
13.1.4 Binary input
The binary input is used for measuring range switching.
Voltage range
0 V to 30 V
Voltage
High
Min.
12 V
Voltage
Low
max.
9.0 V
Current consumption at 24 V
30 mA
Undefined voltage range
9.0 to 12 V
13.2 Output
13.2.1 Output signal
Conductivity:
0 / 4 to 20 mA, galvanically isolated
Temperature:
0 / 4 to 20 mA, galvanically isolated
13.2.2 Load
Max. 500 Ω
13.2.3 Characteristic
Linear
13.2.4 Signal resolution
Resolution:
> 13 bit
Accuracy:
± 20 μA
13.2.5 Alarm output
The alarm output is implemented as an "open collector".
Max. current
200 mA
Max. voltage
30 V DC
Error or device without supply voltage
Alarm output blocked (0 mA)
No error
Alarm output open (up to 200 mA)