Commissioning
Smartec CLD18
28
Hauser
Parameter
Possible settings
Description
MRS
Setting for measuring range switching → 31
Out1 0/4 mA
0 to 2000000 μS/cm
0 μS/cm
‣
Enter the measured value at which the min. current value
(0/4 mA) is present at the transmitter output.
Out1 20 mA
0 to 2000000 μS/cm
0 μS/cm
‣
Enter the measured value at which the max. current value
(20 mA) is present at the transmitter output.
Out2 0/4 mA
-50 to 250 °C
0.0 °C
‣
Enter the measured value at which the min. current value
(0/4 mA) is present at the transmitter output.
Out2 20 mA
-50 to 250 °C
100.0 °C
‣
Enter the measured value at which the max. current value
(20 mA) is present at the transmitter output.
Damping main
0 ... 60 s
0 s
Setting for the damping
Alpha coeff.
1.0 to 20 %/K
2.1 %/K
Coefficient for linear temperature compensation
Factory default
Factory settings
Please confirm
No
No
,
Yes