Smartec CLD18
Diagnostics and troubleshooting
Hauser
35
9.3
Pending diagnostic messages
The diagnostic message consists of a diagnostic code and a message text. The diagnostic code
consists of the error category as per Namur NE 107 and the message number.
Error category (letter in front of the message number):
• F =
Failure
, a malfunction has been detected
The measured value of the affected channel is no longer reliable. Look for the cause in the
measuring point. If a control system is connected, it must be switched to manual mode.
• M =
Maintenance required
, action should be taken as soon as possible
The device still measures correctly. Immediate measures are not necessary. Proper
maintenance efforts may prevent a possible malfunction in the future.
• C =
Function check
, waiting (no error)
Maintenance work is being performed on the device. Wait until the work has been
completed.
• S =
Out of specification
, the measuring point is being operated outside your specification
Operation is still possible. However, you run the risk of increased wear, shorter operating
life or reduced measurement accuracy. Look for the cause in the measuring point.
Diagnostic code
Message text
Description
F61
Sensor elec.
Sensor electronics defective
Remedy:
Contact the Service Department
F62
Sens. Connect
Sensor connection
Remedy:
Contact the Service Department
F100
Sensor comm.
Sensor not communicating
Possible reasons:
No sensor connection
Remedy:
Contact the Service Department
F130
Sensor supply
Sensor check
No conductivity displayed
Possible reasons:
• Sensor in air
• Sensor defective
Remedy:
• Check sensor installation
• Contact the Service Department
F143
Selftest
Sensor self-test error
Remedy:
Contact the Service Department
F152
No airset
Sensor data
No calibration data available
Remedy:
Perform an air set