Diagnostics and troubleshooting
Smartec CLD18
34
Hauser
9
Diagnostics and troubleshooting
9.1
General troubleshooting
User interface
Cause
Solution
No measured value displayed
No power supply connected
Check the device's power supply.
Power is supplied, device is defective
The device must be replaced.
Diagnostic message is displayed
Diagnostic messages → 35
9.2
Trouble shooting instructions
1. Use the 'E' key to call up the main menu.
2. Navigate through the available menus with the '+' and '–' keys.
3. Press the 'E' key to open the desired menu.
4. Use the option,
Back
which can be found at the bottom of each menu, to move up a
level in the menu structure.
Parameter
Possible settings
Description
Current diag.
Read only
Displays the current diagnostic message
Last diag.
Read only
Displays the last diagnostic message
Diag. logbook
Read only
Displays the last diagnostic messages
Device info
Read only
Displays device information
Sensor info
Read only
Displays sensor information
Simulation
Analog out 1 Off
0 mA, 3.6 mA, 4 mA, 10 mA, 12 mA,
20 mA, 21 mA
Outputs a corresponding value at the "
Analog out 1
"
output.
Analog out 2 Off
0 mA, 3.6 mA, 4 mA, 10 mA, 12 mA,
20 mA, 21 mA
Outputs a corresponding value at the "
Analog out 2
"
output.
Alarm out
Off
Active
Inactive
Reset device